Error 21-624, 22-624, or 23-624: Bad DENSITOMETER Data

8F2919
Page
161 of 326
Messages
xxxxx Supply: Film
Calibration Failure (Alarm
sounds)
Description
Note
UPPER DRAWER (U) = 21. MIDDLE DRAWER (M) = 22, LOWER DRAWER (L) = 23.
This error occurs when the density readings from the calibration print are not monotonic or
the DENSITOMETER did not locate exactly 21 steps in the CALIBRATION WEDGE. A
possible cause is fogged or expired film. The error clears when another calibration is asked
for or when the SUPPLY DRAWER closes (as indicated by the SUPPLY DRAWER
INTERLOCK).
Result of Error
The system cannot print from this SUPPLY DRAWER. The SPOOLER is able to receive and
queue jobs, and print from any other available SUPPLY DRAWER.
User Action
1
Install a new FILM CARTRIDGE.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL