DIAGNOSTICS
30JUL07
2
Do the following:
8F2919
Page
209 of 326
CONNECTORS on the:
SOP BOARD - use the
SERVICE TOOL
DRIVE BELT
EXPOSURE ENTRANCE
CLAMP MOTOR - use the
SERVICE TOOL
DRIVE MOTOR - use the
SERVICE TOOL
PINCH ROLLERS
ETB
3
Install a new OPTICS AY. See the ADJUSTMENTS AND REPLACEMENTS,
and EXPOSURE TRANSPORT
Check:
DRE USB CABLE
PDB
ETB
SOP BOARD
If an existing problem:
Seat the CONNECTORS.
Install a new SENSOR. See the ADJUSTMENTS AND
REPLACEMENTS,
START OF PAGE (SOP) BOARD
(S13) - EXPOSURE TRANSPORT
Install a new BELT. See the ILLUSTRATED PARTS LIST,
EXPOSURE TRANSPORT
Install a new MOTOR. See the ADJUSTMENTS AND
REPLACEMENTS,
MOTOR, ENTRANCE/EXIT CLAMP
(M10, M11) - EXPOSURE TRANSPORT
Install a new MOTOR. See the ADJUSTMENTS AND
REPLACEMENTS,
MOTOR, DRIVE (M12) -
EXPOSURE TRANSPORT
Install a new ROLLER. See the ADJUSTMENTS AND
REPLACEMENTS,
PINCH ROLLER AY, CONTINUOUS
- EXPOSURE TRANSPORT AY
AY, SEGMENTED - EXPOSURE TRANSPORT
Install a new ETB. See the ADJUSTMENTS AND
REPLACEMENTS,
EXPOSURE TRANSPORT BOARD
(ETB) - EXPOSURE TRANSPORT
AY.
Status Messages and Error Codes
AY.
AY.
AY.
AY.
and
PINCH ROLLER
AY.
AY.
OPTICS AY