Error 20-915: Dre Data Transfer Fatal Error Detected - Kodak DryView 6800 Diagnostics Manual

Laser imager
Hide thumbs Also See for DryView 6800:
Table of Contents

Advertisement

DIAGNOSTICS
30JUL07

Error 20-915: DRE Data Transfer Fatal Error Detected

8F2919
Page
116 of 326
Messages
Internal Image Data Transfer
Failed
(Alarm sounds)
Description
This error occurs when the wrong number of pixels is transferred between the DRE and the
DPB in the allotted time frame on 3 successive films. A faulty image is printed on each
failure. Jobs associated with the faulty image will be failed, placed in the undelivered queue,
and rescheduled for delivery. The error is cleared when the IMAGER power is cycled.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs, but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films must be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
2
Install a new DATAPATH USB CABLE from the DRE to the OPTICS MODULE.
3
Install a new HARD DRIVE. See the ADJUSTMENTS AND REPLACEMENTS,
DRIVE - DRE
LOCAL PANEL
AY.
LOG
Image transfer failure after
multiple attempts. Wrong
number of bytes delivered in
the allotted time frame.
Status Messages and Error Codes
DICOM
SEVERITY
ELEC DOWN
0-Fatal
HARD

Advertisement

Table of Contents
loading

Table of Contents