Error 27-606: Attenuator Calibration Failed (Non-Monotonic) - Kodak DryView 6800 Diagnostics Manual

Laser imager
Hide thumbs Also See for DryView 6800:
Table of Contents

Advertisement

DIAGNOSTICS
30JUL07

Error 27-606: ATTENUATOR Calibration Failed (Non-Monotonic)

8F2919
Page
198 of 326
Messages
Optics: Calibration Failed
(Alarm sounds)
Description
This error occurs when the BEAM POWER measurements taken during ATTENUATOR
calibration are non-monotonic. The error clears when a calibration is asked for or power is
cycled.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
Optics Calibration Failed.
Power Monitor readings not
monotonic
Status Messages and Error Codes
DICOM
SEVERITY
ELEC DOWN
0-Fatal

Advertisement

Table of Contents
loading

Table of Contents