DIAGNOSTICS
30JUL07

Error "Severity" Level

8F2919
Page
44 of 326
SEVERITY
0-Fatal
1-Alert
DESCRIPTION
A failure occurred
and the system will
have to be restarted
A failure occurred,
but it should be
temporary and can
be repaired by the
user.
LOCAL PANEL
EXAMPLE
System resource
Yes and an
failure
audible alarm
Hardware
failures for
subsystems
Configuration/
registry failure
Software init
failure
Network will not
start up
Database failure
Film jam
Yes and an
audible alarm
Internal
temperature
errors
Status Messages and Error Codes
RECOVERY
NOTIFY
Power cycle is
necessary. Service
might be necessary.
Cycle power and
call Service if error
returns.
The IMAGER
should be able to
continue after the
user fixes the
problem. If failure is
for 1 DRAWER, the
IMAGER should be
able to continue
operating by using
the other
DRAWERS.