Error 06-411: Bad Image Data - Kodak DryView 6800 Diagnostics Manual

Laser imager
Hide thumbs Also See for DryView 6800:
Table of Contents

Advertisement

DIAGNOSTICS
30JUL07

Error 06-411: Bad Image Data

8F2919
Page
85 of 326
Messages
MIM Core: Image Data Error
(Alarm sounds)
Description
The PAGE BUILDER component has encountered an error. The specific error will be in the
log.
Result of Error
The page will not be processed because it could not be rendered because of an internal
error.
User Action
1.
Delete jobs in the "unprintable jobs queue".
2.
Restart the LASER IMAGER from the LOCAL PANEL.
3.
Send another print job from the image source.
4.
If the error occurs again, call Service.
FE Action
LOCAL PANEL
LOG
1.
Width and-or height less
than zero
2.
Bits beyond pixel depth
not zeroed
Status Messages and Error Codes
DICOM
SEVERITY
NORMAL
1-Alert

Advertisement

Table of Contents
loading

Table of Contents