Error 29-926: DENSITOMETER Fiducial Failed Offset Diagnostics

8F2919
Page
239 of 326
Messages
Densitometer: Internal
Diagnostics Failure
(Alarm sounds)
Description
During a calibration print and during DENSITOMETER diagnostics, a reading from the
FIDUCIAL SENSOR (S17) is taken with the light source turned off. The error is indicated if
the reading out of range. The error is cleared when power is cycled on the IMAGER.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
Densitometer fiducial sensor
offset reading out of range
with light source off
Status Messages and Error Codes
DICOM
SEVERITY
ELEC DOWN
0-Fatal