Summary of Contents for Kodak DryView 6800

  • Page 1 { D i a gnost i c s} { P r o duct i o n} { H eal t h Gr o up} { E xt e r n al A ndI n t e r n al } Publication No. 8F2919 30JUL07 DIAGNOSTICS Confidential Restricted Information for the Kodak DryView 6800 LASER IMAGER Service Code: 1649 Important Qualified personnel must repair this equipment. H210_0800GC © Carestream Health, Inc., 2007...
  • Page 2: Table Of Contents

    DIAGNOSTICS 30JUL07 PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the 8F2919 subject matter gained by Carestream Health, Inc., prior to publication. Page 2 of 326 No patent license is granted by this information. Carestream Health, Inc., reserves the right to change this information without notice, and makes no warranty, express or implied, with respect to this information.
  • Page 3 DIAGNOSTICS 30JUL07 Status Messages and Error Codes ........8F2919 Overview.
  • Page 4 DIAGNOSTICS 30JUL07 Error 06-106: Job Delivery Success ....... 8F2919 Error 06-200: MIM Data Component Error .
  • Page 5 DIAGNOSTICS 30JUL07 Error 21-125, 22-125, or 23-125: FEED ROLLER Nip Failure ..8F2919 Error 21-126, 22-126, or 23-126: Jam at Feeder ..... Page 5 of 326 Error 21-139, 22-139, or 23-139: Unable to Read or Write to RF TAG...
  • Page 6 DIAGNOSTICS 30JUL07 Error 27-651: Exposure Timeout Notification ..... . . 8F2919 Error 27-931: OPTICS Communications Failure ..... Page 6 of 326 Error 27-932: OPTICS Communications...
  • Page 7 DIAGNOSTICS 30JUL07 Error 33-386: EXPOSURE TRANSPORT MOTOR Speed Control Failure 8F2919 Error 33-387: EXPOSURE TRANSPORT MOTOR Speed Control Warning Page 7 of 326 Error 33-931: EXPOSURE TRANSPORT Communications Failure ..Error 33-932: EXPOSURE TRANSPORT Communications Retry .
  • Page 8 DIAGNOSTICS Using the Diagnostics 30JUL07 Section 1: Using the Diagnostics 8F2919 Page 8 of 326 Using the SERVICE SWITCH The SERVICE SWITCH is located inside the MAIN DOOR behind the LATCH MOTOR. SERVICE POSITION Top SCREW BRACKET Bottom SCREW H210_0582AC Open the MAIN DOOR (see the INSTALLATION INSTRUCTIONS, Using the Manual LATCH...
  • Page 9: Using The Diagnostics

    DIAGNOSTICS Using the Diagnostics 30JUL07 SERVICE TOOL 8F2919 Page Functions 9 of 326 The SERVICE TOOL provides the following functions. • Configuration - Provides tools for configuration of the IMAGER. • Diagnostics - Provides tools to test the components of the IMAGER. •...
  • Page 10: Connecting/Disconnecting The Laptop Computer To/From The Imager

    DIAGNOSTICS Using the Diagnostics 30JUL07 Connecting/Disconnecting the LAPTOP COMPUTER to/from the IMAGER 8F2919 Page 10 of 326 AIR INTAKE DOOR RJ45 CONNECTOR ELECTRONICS FILTER 2 SCREWS AIR INTAKE DOOR Caution Dangerous Voltage To connect: Open the AIR INTAKE DOOR. Remove: •...
  • Page 11 DIAGNOSTICS Using the Diagnostics 30JUL07 8F2919 Page 11 of 326 CROSSOVER CABLE AIR INTAKE DOOR To disconnect: Disconnect the CROSSOVER CABLE from the IMAGER. Install: • ELECTRONICS FILTER • 3 NUTS Close the AIR INTAKE DOOR.
  • Page 12: Logging On To The Service Tool

    The main menu of the SERVICE TOOL displays. Logging Off From the SERVICE TOOL In the upper right corner of the SERVICE TOOL, click [Log Off]. At the “Kodak Service WebLink” screen, click [Log Off]. Close the “WebLink” Client Software. At the “Kodak Secure Link 2.1” screen: •...
  • Page 13: Stopping And Starting The Remote Management Services

    DIAGNOSTICS Using the Diagnostics 30JUL07 Stopping and Starting the REMOTE MANAGEMENT SERVICES 8F2919 Page At times it may be necessary to stop, start, or restart the REMOTE MANAGEMENT 13 of 326 SERVICES. Connect the LAPTOP COMPUTER to the IMAGER (see Connecting/Disconnecting the LAPTOP COMPUTER to/from the IMAGER).
  • Page 14: Diagnostic Tests

    DIAGNOSTICS Using the Diagnostics 30JUL07 • stop the agent, click [Stop] 8F2919 Page • start the agent, click [Start] 14 of 326 • restart the agent, click [Restart] Click [Save]. Diagnostic Tests DENSITOMETER Select Diagnostics>MCS>Densitometer. Click [Run]. Note This is the same test run at power ON for the DENSITOMETER. The test ends with a “Pass” or “Fail”...
  • Page 15: Motors

    DIAGNOSTICS Using the Diagnostics 30JUL07 MOTORS 8F2919 Page 15 of 326 Important This test is not for the ROLLBACK AY or PICKUP AY. To test these MOTORS see TRAY Diagnostics. Select Diagnostics>MCS>Motors Diagnostics. Set the SERVICE SWITCH in the service position. From the “Select Operation”...
  • Page 16: Mcs Microcontrollers

    DIAGNOSTICS Using the Diagnostics 30JUL07 MCS MICROCONTROLLERS 8F2919 Page 16 of 326 Important This test is not for the ROLLBACK AY or PICKUP AY. To test these MICROCONTROLLERS TRAY Diagnostics. To test any of the MICROCONTROLLERS on the I C bus, select Diagnostics>MCS>MCS Diagnostics.
  • Page 17: Tray Diagnostics

    DIAGNOSTICS Using the Diagnostics 30JUL07 TRAY Diagnostics 8F2919 Page Select Diagnostics>MCS>Tray Diagnostics. 17 of 326 Set the SERVICE SWITCH in the service position. From the “Select Diagnostic” drop-down menu, select “Upper Tray,” “Middle Tray,” or “Lower Tray.” From the “Select Operation” drop-down menu, select the type of test. The test types include: •...
  • Page 18 DIAGNOSTICS Using the Diagnostics 30JUL07 From the “Cycle” drop-down menu, select the test option, “One Cycle” or “Continuous”. 8F2919 Page 18 of 326 Note If “Continuous” is selected, the test will run until “Stop” is selected. Continuous should only be used for test and film transport to a sorter bin.
  • Page 19: Optics

    DIAGNOSTICS Using the Diagnostics 30JUL07 OPTICS 8F2919 Page 19 of 326 Starting the Tests Select Diagnostics>MCS>Optics. OPTICS Calibration This calibration of the OPTICS is the same that occurs before a calibration print or when the IMAGER is energized. The calibration: •...
  • Page 20 DIAGNOSTICS Using the Diagnostics 30JUL07 From the Main Menu of the SERVICE TOOL, select Diagnostics>Diag Summary. 8F2919 Page From the diagnostics “Summary” screen, select “Test Results.” 20 of 326 From the “Diagnostic” drop-down menu, select “SDSFilmCalibDiagnostics - getFilmCalibData.” Click [Save]. Store the data in a folder on the HARD DRIVE of your LAPTOP COMPUTER.
  • Page 21: Sensors

    DIAGNOSTICS Using the Diagnostics 30JUL07 SENSORS 8F2919 Page Select Monitoring and Control>Sensors Monitor. 21 of 326 Select the check box of the SENSOR to be monitored. Note Up to 3 SENSORS can be monitored. Click [Start]. To stop monitoring, click [Stop].
  • Page 22 DIAGNOSTICS Using the Diagnostics 30JUL07 To “pause” monitoring: 8F2919 Page • Click [Pause] to stop. 22 of 326 • Click [Resume] to continue. To monitor the SENSORS when film is moving through the system: From the Main Menu of the SERVICE TOOL, select Diagnostics>DRE>Test Print. Select an image to print from the drop-down menu.
  • Page 23: Sorter

    DIAGNOSTICS Using the Diagnostics 30JUL07 SORTER 8F2919 Page The SERVICE TOOL includes 2 diagnostics for the SORTER: 23 of 326 SORTER SENSOR Calibration - This is the MCS diagnostic run normally at power on. The test, run from the top part of the screen, displays diagnostic data, and indicates OK (or “pass”).
  • Page 24 DIAGNOSTICS Using the Diagnostics 30JUL07 To check operation of the SORTER SENSOR, select Diagnostics>MCS>Sorter. 8F2919 Page To do calibration of the SORTER SENSOR: 24 of 326 Set the SERVICE SWITCH in the “Service” mode. In the top right corner of the screen, click [Run]. The calibration will run and indicate OK.
  • Page 25: Film Cartridge Information

    DIAGNOSTICS Using the Diagnostics 30JUL07 Film Cartridge Information 8F2919 Page To obtain information about the film loaded in any DRAWER, select Hardware Data>Film 25 of 326 Cartridge Data. From the “Select Tray” drop-down menu, select the DRAWER from which you want film information: “Upper,”...
  • Page 26: Film Calibration Data

    DIAGNOSTICS Using the Diagnostics 30JUL07 Film Calibration Data 8F2919 Page To obtain a “DLogE Curve” from the last calibration data for film in any CARTRIDGE in 26 of 326 the IMAGER, select Hardware Data>Film Calibration. Select the SUPPLY DRAWER you want to check. Click: •...
  • Page 27: Diagnostics Summary Information

    DIAGNOSTICS Using the Diagnostics 30JUL07 Diagnostics Summary Information 8F2919 Page Select Diagnostics>Diag Summary. 27 of 326 To view the status of any diagnostic that is now running, select “Current Status.” If a diagnostics test is running, the name of the test and the “Insurance” Number of the user will display.
  • Page 28 DIAGNOSTICS Using the Diagnostics 30JUL07 8F2919 Page 28 of 326 From the “Diagnostic” drop-down menu, select the test you want. To display a “Summary” of the diagnostic, click [View]. If you want to copy the test summary to the HARD DRIVE of your LAPTOP, click [Save]. Store the file in a folder on the HARD DRIVE of your LAPTOP COMPUTER.
  • Page 29: Local Panel

    DIAGNOSTICS Using the Diagnostics 30JUL07 LOCAL PANEL 8F2919 Page Select Diagnostics>MCS>Local Panel. 29 of 326 Click [Run]. The SPEAKER in the LOCAL PANEL should beep. This indicates: • The DRE can communicate with the LOCAL PANEL. • The SPEAKER in the LOCAL PANEL operates.
  • Page 30: Communication On The Network

    DIAGNOSTICS Using the Diagnostics 30JUL07 Communication on the Network 8F2919 Page To test the connection of the IMAGER with the network, select Diagnostics>Network. 30 of 326 To do a “Ping” test to check that the IMAGER can contact other devices on the network: Enter the IP Address of a device on the network.
  • Page 31 DIAGNOSTICS Using the Diagnostics 30JUL07 To run a DICOM “Echo” test to check if the IMAGER can do communication with a 8F2919 Page another device on the network, select the “DICOM Echo” tab. 31 of 326 For the device, type: •...
  • Page 32: Temperature Of The Processor

    DIAGNOSTICS Using the Diagnostics 30JUL07 Temperature of the PROCESSOR 8F2919 Page To check the temperatures of the PROCESSOR “zones”, select Monitoring and 32 of 326 Control>Temperature Monitor.
  • Page 33: Dre Test Print

    DIAGNOSTICS Using the Diagnostics 30JUL07 DRE Test Print 8F2919 Page To select and run a test print stored in the DRE, select Diagnostics>DRE>Test Print. 33 of 326 Select the parameters for the DRE Test Print. In the “Study Label” window, enter a name (up to 20 characters) for the print. From the “Test Print”...
  • Page 34: Power-On "Self-Tests

    DIAGNOSTICS Using the Diagnostics 30JUL07 Power-on “Self-Tests” 8F2919 Page “Self-test” and calibration are automatically run by the system when power is applied. If an 34 of 326 error occurs, an error code is displayed on the LOCAL PANEL. Under control of the MCS MASTER CPU, tests and calibration are run by MICROCONTROLLERS that control the other BOARDS on the I C bus of the MCS.
  • Page 35: Jam Areas

    DIAGNOSTICS Using the Diagnostics 30JUL07 Jam Areas 8F2919 Page The following are the areas indicated by the film jam errors. 35 of 326 Area 5 Area 4 Area 3 Area 1U Area 1M Area 1L Area 2 = Knob...
  • Page 36: 8F2919 Status Messages And Error Codes

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Section 2: Status Messages and Error Codes 8F2919 Page 36 of 326 Overview This section provides information about 5 types of system messages: • DICOM status messages from the LASER IMAGER to the SCU •...
  • Page 37 DIAGNOSTICS Status Messages and Error Codes 30JUL07 8F2919 DICOM PRINTER DICOM PRINTER Status Information Page 37 of 326 Status Message Message FAILURE Condition PRINTER DOWN ELEC DOWN PROC DOWN WARNING Condition BAD SUPPLY MGZ CALIBRATION ERR CHECK PRINTER COVER OPEN EMPTY MEDIASZ MEDIATP FILM JAM FILM TRANS ERR...
  • Page 38: Printer Status Messages

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 PRINTER Status Messages 8F2919 Page The LOCAL PANEL displays information to the user describing the status of the PRINTER. 38 of 326 DICOM PRINTER LOCAL PANEL PRINTER STATUS Notes STATUS/INFO Close all doors to DOOR open WARNING/COVER Any of the IMAGER...
  • Page 39 DIAGNOSTICS Status Messages and Error Codes 30JUL07 8F2919 DICOM PRINTER LOCAL PANEL PRINTER STATUS Notes Page STATUS/INFO 39 of 326 Service Mode Service Mode WARNING/ The SERVICE PRINTER OFFLINE OVERRIDE SWITCH is enabled. The IMAGER is not ready. Shutdown Initiated Shutdown Initiated FAILURE/PRINTER When shutdown is...
  • Page 40 DIAGNOSTICS Status Messages and Error Codes 30JUL07 8F2919 DICOM PRINTER LOCAL PANEL PRINTER STATUS Notes Page STATUS/INFO 40 of 326 Unlocking, please IMAGER unlock The PRINTER is wait requested completing prints in process. Warming= Warming WARNING/PROC The PROCESSOR is INIT “warming”...
  • Page 41: Film Tray Status Messages

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 FILM TRAY Status Messages 8F2919 Page The LOCAL PANEL displays information to the user describing the status of the FILM TRAYS. 41 of 326 LOCAL PANEL FILM SUPPLY STATE Notes Calibrating Calibrating A calibration is in process for this FILM CARTRIDGE.
  • Page 42: Job Manager Status Messages

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 JOB MANAGER Status Messages 8F2919 Page The LOCAL PANEL displays information to the user describing the status of the jobs. 42 of 326 LOCAL PANEL JOB MANAGER Status Notes Message Imager Offline Offline The IMAGER will not “accept”...
  • Page 43 DIAGNOSTICS Status Messages and Error Codes 30JUL07 8F2919 Subsystem ID Subsystem/Component Name Error Code “Range” Page 43 of 326 MIM CORE Components: “01-xxx - 09-xxx” - CONFIGURATION and CONTROL SUBSYSTEM - MESSAGING SUBSYSTEM - STATISTICS SUBSYSTEM - DICOM INPUT PRINT SCP SUBSYSTEM - STUDY RECLAMATION SUBSYSTEM - DELIVERY SUBSYSTEM - MIM DATA COMPONENTS SUBSYSTEM...
  • Page 44: Error "Severity" Level

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error “Severity” Level 8F2919 Page 44 of 326 LOCAL PANEL SEVERITY DESCRIPTION EXAMPLE RECOVERY NOTIFY 0-Fatal A failure occurred • System resource Yes and an Power cycle is and the system will failure audible alarm necessary.
  • Page 45 DIAGNOSTICS Status Messages and Error Codes 30JUL07 8F2919 LOCAL PANEL SEVERITY DESCRIPTION EXAMPLE RECOVERY Page NOTIFY 45 of 326 2-Warning An “event” occurred • Preventive Yes, but no No recovery that should notify the maintenance audible alarm necessary. user. 3-Notify An “event”...
  • Page 46: Error 01-000: General Configuration And Control Subsystem

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 01-000: General CONFIGURATION AND CONTROL SUBSYSTEM 8F2919 Page Error 46 of 326 Messages LOCAL PANEL DICOM SEVERITY None General Error in NORMAL 3-Notify (No alarm) Configuration and Control Subsystem Description A general, non-specific error has occurred in the CONFIGURATION AND CONTROL SUBSYSTEM.
  • Page 47: Error 01-001: Maximum Number Of Restarts Exceeded

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 01-001: Maximum Number of Restarts Exceeded 8F2919 Page 47 of 326 Messages LOCAL PANEL DICOM SEVERITY None Maximum Number of NORMAL 2-Warning (No alarm) Restarts Exceeded for xxxx (xxxx is the subsystem or component name) Description When a subsystem or component generates a non-recoverable, CONFIGURATION AND...
  • Page 48: Error 01-002: Error Reading Configuration Information

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 01-002: Error Reading Configuration Information 8F2919 Page 48 of 326 Messages LOCAL PANEL DICOM SEVERITY None Error Reading Configuration NORMAL 2-Warning (No alarm) Information Description An error occurred when reading the CONFIGURATION AND CONTROL SUBSYSTEM configuration information.
  • Page 49: Error 01-004: Error Starting A Subsystem

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 01-004: Error Starting a Subsystem 8F2919 Page 49 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Internal Software Error starting subsystem xxxx NORMAL 1-Alert Error (xxxx is the subsystem name) (Alarm sounds) Description An error occurred when attempting to start a subsystem.
  • Page 50: Error 01-005: Error Stopping A Subsystem

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 01-005: Error Stopping a Subsystem 8F2919 Page 50 of 326 Messages LOCAL PANEL DICOM SEVERITY None Error stopping subsystem NORMAL 1-Alert (No alarm) xxxx (xxxx is the subsystem name) Description An error occurred when attempting to stop a subsystem. Result of Error The subsystem in question might be running.
  • Page 51: Error 01-007: General Subsystem Communication Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 01-007: General Subsystem Communication Error 8F2919 Page 51 of 326 Messages LOCAL PANEL DICOM SEVERITY None Configuration and Control NORMAL 1-Alert (No alarm) Subsystem cannot communicate with subsystem xxxx (xxxx is the subsystem name) Description An unknown error occurred when communicating with a subsystem.
  • Page 52: Error 01-008: All Subsystems Started

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 01-008: All Subsystems Started 8F2919 Page 52 of 326 Messages LOCAL PANEL DICOM SEVERITY None All subsystems successfully NORMAL 3-Notify (No alarm) started Description All subsystems to be started by the CONFIGURATION AND CONTROL SUBSYSTEM have started.
  • Page 53: Error 01-009: Operating System Could Not Be Shut Down

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 01-009: OPERATING SYSTEM Could Not Be Shut Down 8F2919 Page 53 of 326 Messages LOCAL PANEL DICOM SEVERITY None OS could not be shut down NORMAL 3-Notify (No alarm) Description The OPERATING SYSTEM did not complete shutting down in the expected time. Result of Error None, the system will function normally.
  • Page 54: Error 01-010: Subsystem Terminated Abnormally

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 01-010: Subsystem Terminated Abnormally 8F2919 Page 54 of 326 Messages LOCAL PANEL DICOM SEVERITY None Subsystem xxxx terminated NORMAL 3-Notify (No alarm) (xxxx is the subsystem name) Description A subsystem that was started by the CONFIGURATION AND CONTROL SUBSYSTEM terminated abnormally.
  • Page 55: Error 04-000: General Dicom Input Print Scp Subsystem Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 04-000: General DICOM INPUT PRINT SCP SUBSYSTEM Error 8F2919 Page 55 of 326 Messages LOCAL PANEL DICOM SEVERITY None General Error in DICOM Input NORMAL 3-Notify (No alarm) Print SCP Subsystem Description A general, non-specific error has occurred in the DICOM INPUT PRINT SCP SUBSYSTEM.
  • Page 56: Error 04-100: Limit On Number Of Associations Exceeded

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 04-100: Limit on Number of Associations Exceeded 8F2919 Page 56 of 326 Messages LOCAL PANEL DICOM SEVERITY None Maximum Number of NORMAL 1-Alert (No alarm) Associations Exceeded Description The maximum number of associations is exceeded. The association is rejected. Result of Error The SCU will not be able to open an association, and will not be able to deliver a print.
  • Page 57: Error 04-101: Ae Title Does Not Correspond To A Recognized Printer

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 04-101: AE Title does not Correspond to a Recognized PRINTER 8F2919 Page 57 of 326 Messages LOCAL PANEL DICOM SEVERITY None No Destination Matching NORMAL 1-Alert (No alarm) Called AE Title Description The AE title did not map to a destination that is available at this SCP.
  • Page 58: Error 04-110: Invalid Association Request

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 04-110: Invalid Association Request 8F2919 Page 58 of 326 Messages LOCAL PANEL DICOM SEVERITY None The association request was NORMAL 1-Alert (No alarm) rejected Description An association request was rejected for one or more of the following: •...
  • Page 59: Error 04-200: Disk Full

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 04-200: DISK Full 8F2919 Page 59 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Disk Full Disk Full NORMAL 1-Alert (Alarm sounds) Description The area of the DISK in which studies are stored is full. Result of Error If a study is in progress, it cannot be completed.
  • Page 60: Error 04-201: Keep-Alive Message Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 04-201: Keep-Alive Message Failure 8F2919 Page 60 of 326 Messages LOCAL PANEL DICOM SEVERITY None Send Keep-Alive Message NORMAL 2-Warning (No alarm) Failed Description The SCP might be configured to send packets to maintain a TCP/IP connection with an SCU, to which the SCU should reply.
  • Page 61: Error 04-202: Connection Is Lost

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 04-202: Connection Is Lost 8F2919 Page 61 of 326 Messages LOCAL PANEL DICOM SEVERITY None Network lost connection NORMAL 2-Warning (No alarm) Description The SCP has lost the network connection to an SCU. Result of Error The SCP will abort the association tied to this connection.
  • Page 62: Error 04-300: Plut Is Missing

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 04-300: PLUT is Missing 8F2919 Page 62 of 326 Messages LOCAL PANEL DICOM SEVERITY None Referencing PLUT Missing NORMAL 1-Alert (No alarm) Description The PLUT indicated by tag (2050,0500) in the FILM BOX or IMAGE BOX is not present. Result of Error The MIM CORE will not process the print job.
  • Page 63: Error 04-301: Missing Necessary Attribute

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 04-301: Missing Necessary Attribute 8F2919 Page 63 of 326 Messages LOCAL PANEL DICOM SEVERITY None Required attribute wwww NORMAL 1-Alert (No alarm) (xxxx,yyyy) is missing from zzzz SOP (wwww (xxxx,yyyy) is the tag name and number; zzzz is the SOP name) Description A necessary attribute for the SOP is missing.
  • Page 64: Error 04-302: Invalid Attribute

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 04-302: Invalid Attribute 8F2919 Page 64 of 326 Messages LOCAL PANEL DICOM SEVERITY None Invalid attribute wwww NORMAL 1-Alert (No alarm) (xxxx,yyyy) encountered in zzzz SOP (wwww (xxxx,yyyy) is the tag name and number; zzzz is the SOP name) Description An invalid attribute for the SOP was found.
  • Page 65: Error 05-000: Error Unknown

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 05-000: Error Unknown 8F2919 Page 65 of 326 Messages LOCAL PANEL DICOM SEVERITY None Error Unknown NORMAL 2-Warning (No alarm) Description An unknown error has occurred in the STUDY RECLAMATION SUBSYSTEM. Result of Error Unknown.
  • Page 66: Error 05-001: Error Reading Configuration Information

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 05-001: Error Reading Configuration Information 8F2919 Page 66 of 326 Messages LOCAL PANEL DICOM SEVERITY None Error reading configuration NORMAL 2-Warning (No alarm) information Description An error occurred reading the STUDY RECLAMATION SUBSYSTEM configuration information.
  • Page 67: Error 05-002: Study Reclamation Initiated

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 05-002: Study Reclamation Initiated 8F2919 Page 67 of 326 Messages LOCAL PANEL DICOM SEVERITY None High water-mark reached. NORMAL 3-Notify (No alarm) Scheduled study reclamation time reached Description Study reclamation is initiated because either the high-water mark is reached, or the scheduled time for study reclamation has occurred.
  • Page 68: Error 05-003: Low-Water Mark Cannot Be Reached

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 05-003: Low-Water Mark Cannot be Reached 8F2919 Page 68 of 326 Messages LOCAL PANEL DICOM SEVERITY None Low water-mark could not be NORMAL 3-Notify (No alarm) reached Description The STORAGE RECOVERY SUBSYSTEM could not reach the low-water mark when running the STORAGE RECLAMATION ROUTINES.
  • Page 69: Error 05-004: Low-Water Mark Reached

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 05-004: Low-Water Mark Reached 8F2919 Page 69 of 326 Messages LOCAL PANEL DICOM SEVERITY None Low-water-mark reached NORMAL 3-Notify (No alarm) Description Study reclamation has is completed because the low-water mark was successfully reached. Result of Error None User Action...
  • Page 70: Error 05-005: Study Reclamation Shutting Down

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 05-005: Study Reclamation Shutting Down 8F2919 Page 70 of 326 Messages LOCAL PANEL DICOM SEVERITY None Study reclamation is shutting NORMAL 3-Notify (No alarm) down Description The STUDY RECLAMATION SUBSYSTEM has received a shutdown request. Result of Error None.
  • Page 71: Error 06-000: General Delivery Subsystem Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-000: General DELIVERY SUBSYSTEM Error 8F2919 Page 71 of 326 Messages LOCAL PANEL DICOM SEVERITY None General error in delivery NORMAL 3-Notify (No alarm) subsystem Description A general, non-specific error has occurred in the DELIVERY SUBSYSTEM. Result of Error Unknown.
  • Page 72: Error 06-001: General Start Up Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-001: General Start UP Error 8F2919 Page 72 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Internal Software General start up error NORMAL 1-Alert Error (Alarm sounds) Description A general error occurred starting the DELIVERY SUBSYSTEM. Result of Error Jobs will not be completed, as the software interface for the DELIVERY SUBSYSTEM could not be started.
  • Page 73: Error 06-002: Bad Or Missing Configuration Data

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-002: Bad or Missing Configuration Data 8F2919 Page 73 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Internal Software Bad or missing configuration NORMAL 1-Alert Error information (Alarm sounds) Description An error occurred reading the DELIVERY SUBSYSTEM configuration information. Result of Error The DELIVERY SUBSYSTEM might not execute correctly.
  • Page 74: Error 06-101: Internal Job Processing Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-101: Internal Job Processing Error 8F2919 Page 74 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Internal Software General job processing error. NORMAL 1-Alert Error Start up error (Alarm sounds) Bad or missing configuration data Resource allocation error Mode switching error...
  • Page 75: Error 06-102: Printer Not Ready

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-102: PRINTER Not Ready 8F2919 Page 75 of 326 Messages LOCAL PANEL DICOM SEVERITY None Printer not in ready state NORMAL 2-Warning (No alarm) Description The PRINTER is not ready to receive delivery jobs. Result of Error Jobs will not be completed when the PRINTER is in this state.
  • Page 76: Error 06-103: Invalid Media

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-103: Invalid Media 8F2919 Page 76 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Internal Software Invalid media NORMAL 1-Alert Error (Alarm sounds) Description An error occurred reading the DELIVERY SUBSYSTEM configuration information. Result of Error The job will not be completed.
  • Page 77: Error 06-104: Media Not Available

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-104: Media Not Available 8F2919 Page 77 of 326 Messages LOCAL PANEL DICOM SEVERITY None Media unavailable NORMAL 2-Warning (No alarm) Description The media is not available. Result of Error The job will not be completed. User Action None FE Action...
  • Page 78: Error 06-105: Page Delivery Success

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-105: Page Delivery Success 8F2919 Page 78 of 326 Messages LOCAL PANEL DICOM SEVERITY None Page delivery success NORMAL 3-Notify (No alarm) Description A page was successfully completed. This is not an error condition. Result of Error None User Action...
  • Page 79: Error 06-106: Job Delivery Success

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-106: Job Delivery Success 8F2919 Page 79 of 326 Messages LOCAL PANEL DICOM SEVERITY None Job delivery success NORMAL 3-Notify (No alarm) Description A job was successfully completed. This is not an error condition. Result of Error None User Action...
  • Page 80: Error 06-200: Mim Data Component Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-200: MIM Data Component Error 8F2919 Page 80 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Internal Software Database access error NORMAL 1-Alert Error (Alarm sounds) Description An error occurred when trying to access the MIM CORE DATABASE. The DATABASE might be corrupted, or a timeout occurred.
  • Page 81: Error 06-300: Destination Business Logic Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-300: Destination Business Logic Error 8F2919 Page 81 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Internal Software Database access error NORMAL 1-Alert Error (Alarm sounds) Description An error occurred in the DESTINATION BUSINESS LOGIC component. The DATABASE might be corrupted, or a timeout occurred.
  • Page 82: Error 06-400: General Page Builder Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-400: General Page Builder Error 8F2919 Page 82 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Image Page Error General page builder error NORMAL 1-Alert (Alarm sounds) Description A general page builder error has occurred. Conditions include: •...
  • Page 83: Error 06-410: Rendering Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-410: Rendering Error 8F2919 Page 83 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Image Rendering General error rendering NORMAL 1-Alert Error the page (Alarm sounds) Error after completion of chain-forming Error rendering trim or rectangle Error rendering text: xxxx...
  • Page 84 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Send another print job from the image source. 8F2919 Page If the error occurs again, call Service. 84 of 326 FE Action...
  • Page 85: Error 06-411: Bad Image Data

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-411: Bad Image Data 8F2919 Page 85 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Image Data Error Width and-or height less NORMAL 1-Alert (Alarm sounds) than zero Bits beyond pixel depth not zeroed Description The PAGE BUILDER component has encountered an error.
  • Page 86: Error 06-412: Abort Requested

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-412: Abort Requested 8F2919 Page 86 of 326 Messages LOCAL PANEL DICOM SEVERITY None Render abort requested NORMAL 1-Alert (No alarm) Description The PAGE BUILDER component is asked to stop a rendering that is in progress. Result of Error The page will not be processed because rendering was stopped.
  • Page 87: Error 06-420: Invalid Data Or Data Access Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-420: Invalid Data or Data Access Error 8F2919 Page 87 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Internal Software Page width and/or height NORMAL 1-Alert Error are less than zero (Alarm sounds) Invalid page band Invalid coordinates for...
  • Page 88 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Result of Error 8F2919 Page The page will not be processed because it could not be rendered because of an internal 88 of 326 error. User Action Delete jobs in the “unprintable jobs queue”. Restart the LASER IMAGER from the LOCAL PANEL.
  • Page 89: Error 06-430: Chain-Building Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-430: Chain-Building Error 8F2919 Page 89 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Internal Software General error building NORMAL 1-Alert Error image processing chain (Alarm sounds) Error in tonescaling: xxxx (xxxx is the specific tonescaling problem) Error forming the color...
  • Page 90: Error 06-500: Internal Page Delivery Component Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-500: Internal PAGE DELIVERY Component Error 8F2919 Page 90 of 326 Messages LOCAL PANEL DICOM SEVERITY None BIND error. Code = xx; NORMAL 1-Alert (No alarm) Page ID = yyy STAGE error. Code = xx; Page ID = yyy PRINT error.
  • Page 91: Error 06-501: Media Not Loaded

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-501: Media Not Loaded 8F2919 Page 91 of 326 Messages LOCAL PANEL DICOM SEVERITY None Media xxxx not loaded (xxxx NORMAL 2-Warning (No alarm) is the requested media) Description The requested media is not loaded in the PRINTER. Result of Error The job will not be completed until the media is loaded.
  • Page 92: Error 06-502: Media Needs Calibration

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-502: Media Needs Calibration 8F2919 Page 92 of 326 Messages LOCAL PANEL DICOM SEVERITY None Media xxxx requires NORMAL 2-Warning (No alarm) calibration (xxxx is the requested media) Description The media is not calibrated. Result of Error The job will not be completed until the media is calibrated.
  • Page 93: Error 06-503: Media Is Calibrating

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-503: Media is Calibrating 8F2919 Page 93 of 326 Messages LOCAL PANEL DICOM SEVERITY None Media xxxx is calibrating NORMAL 3-Notify (No alarm) (xxxx is the requested media) Description The media is calibrating. Result of Error Jobs will not be completed until the calibration cycle is complete.
  • Page 94: Error 06-504: Calibration Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-504: Calibration Error 8F2919 Page 94 of 326 Messages LOCAL PANEL DICOM SEVERITY None Error calibrating media xxxx NORMAL 3-Notify (No alarm) (xxxx is the media that failed calibration) Description An error occurred during media calibration. Result of Error Jobs might not be completed because the media could not be calibrated.
  • Page 95: Error 06-505: Media Is Empty

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-505: Media is Empty 8F2919 Page 95 of 326 Messages LOCAL PANEL DICOM SEVERITY None Media xxxx is empty (xxxx is NORMAL 2-Warning (No alarm) the requested media) Description The media is empty. Result of Error The job will not be completed until the media is loaded.
  • Page 96: Error 06-506: Jam In Media

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-506: Jam in Media 8F2919 Page 96 of 326 Messages LOCAL PANEL DICOM SEVERITY None Media xxxx is jammed (xxxx NORMAL 2-Warning (No alarm) is the requested media) Description There is a jam in the media. Result of Error The job will not be completed until the media jam is cleared.
  • Page 97: Error 06-600: General Delivery Engine Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 06-600: General Delivery Engine Error 8F2919 Page 97 of 326 Messages LOCAL PANEL DICOM SEVERITY MIM Core: Internal Software General error in delivery NORMAL 1-Alert Error subsystem (Alarm sounds) Description A general, non-specific error has occurred in the DELIVERY SUBSYSTEM. Result of Error Unknown.
  • Page 98: Error 10-001: Internal Software Error In Mis

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 10-001: Internal Software Error in MIS 8F2919 Page 98 of 326 Messages LOCAL PANEL DICOM SEVERITY MIS: Internal Software Error MIS internal error PRINTER DOWN 0-Fatal (Alarm sounds) Description This indicates any of a number of errors that indicate that internal software functionality of the MCS INTERFACE SERVER (MIS) SUBSYSTEM has failed.
  • Page 99: Error 10-003: Image Buffer Error In Mis

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 10-003: Image Buffer Error in MIS 8F2919 Page 99 of 326 Messages LOCAL PANEL DICOM SEVERITY MIS: Image Buffer Error MIS image buffer error PRINTER DOWN 0-Fatal (Alarm sounds) Description This indicates any of a number of errors that indicate that image memory buffer functionality of the MCS INTERFACE SERVER (MIS) SUBSYSTEM has failed.
  • Page 100: Error 10-015: Database Error In Mis

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 10-015: Database Error in MIS 8F2919 Page 100 of 326 Messages LOCAL PANEL DICOM SEVERITY MIS: Database Error MIS Database Error CHECK 1-Alert (Alarm sounds) PRINTER Description This indicates any of a number of errors indicating that database software functionality of the MCS INTERFACE SERVER (MIS) SUBSYSTEM has failed.
  • Page 101: Error 10-910: No Communications With The Mcs

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 10-910: No Communications with the MCS 8F2919 Page 101 of 326 Messages LOCAL PANEL DICOM SEVERITY MIS: MCS Communications MIS to MCS Communications PRINTER DOWN 0-Fatal Failure (Alarm sounds) Fail Description This error occurs when the MCS does not respond to the MIS. This error is a fatal error that will result in a service call.
  • Page 102: Error 12-075: Local Panel Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 12-075: LOCAL PANEL Error 8F2919 Page 102 of 326 Messages LOCAL PANEL DICOM SEVERITY None Local Panel Internal Error NORMAL 1-Alert (No alarm) Description General internal error detected in the LOCAL PANEL SUBSYSTEM. The log message will have additional details.
  • Page 103: Error 12-080: Software Update Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 12-080: Software Update Error 8F2919 Page 103 of 326 Messages LOCAL PANEL DICOM SEVERITY None Software Update Failure NORMAL 1-Alert (No alarm) Description The LOCAL PANEL cover screen was notified that the software update was not successfully completed.
  • Page 104: Error 12-085: Local Panel Flash Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 12-085: LOCAL PANEL Flash Error 8F2919 Page 104 of 326 Messages LOCAL PANEL DICOM SEVERITY None Local Panel - Flash Access NORMAL 1-Alert (No alarm) Error Description This error occurs when a problem with the FLASH device (hardware or DRIVER) is encountered when the LOCAL PANEL is attempting to use it.
  • Page 105: Error 13-001: Service Tool Internal Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 13-001: SERVICE TOOL Internal Error 8F2919 Page 105 of 326 Messages LOCAL PANEL DICOM SEVERITY None General Error in Service Tool NORMAL 3-Notify (No alarm) Subsystem Description A SERVICE TOOL internal error has occurred. This indicates any of a number of errors indicating that internal software functionality of the SERVICE TOOL SUBSYSTEM has failed.
  • Page 106: Error 13-002: Service Tool Database Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 13-002: SERVICE TOOL Database Error 8F2919 Page 106 of 326 Messages LOCAL PANEL DICOM SEVERITY None Service Tool Database Error NORMAL 3-Notify (No alarm) Description This indicates any of a number of errors indicating that database software functionality of the SERVICE TOOL SUBSYSTEM has failed.
  • Page 107: Error 20-004: Usb Did Not Initialize

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-004: USB Did Not initialize 8F2919 Page 107 of 326 Messages LOCAL PANEL DICOM SEVERITY USB failure USB failed to initialize PRINTER DOWN 0-Fatal (Alarm sounds) Description This error occurs when the DRE cannot successfully initialize the USB circuits and cannot transfer image data.
  • Page 108: Error 20-154: Mcs Hard Drive Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-154: MCS HARD DRIVE Failure 8F2919 Page 108 of 326 Messages LOCAL PANEL DICOM SEVERITY MCS: Internal MCS failed to read/write to ELEC DOWN 0-Fatal Communications Failure the hard drive (Alarm sounds) Description This error occurs when the MCS MICROCONTROLLER detect an error when reading or writing from NVRAM.
  • Page 109: Error 20-156: Incompatible Software Versions Installed

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-156: Incompatible Software Versions Installed 8F2919 Page 109 of 326 Messages LOCAL PANEL DICOM SEVERITY Incompatible Software One or more micros have PRINTER DOWN 0-Fatal Versions Installed software versions installed (Alarm sounds) that are incompatible with the MCS version Description...
  • Page 110 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Install a new MICRO BOARD. 8F2919 Page Install a new HARD DRIVE. See the ADJUSTMENTS AND REPLACEMENTS, HARD 110 of 326 DRIVE - DRE Install a new DRE. See the ADJUSTMENTS AND REPLACEMENTS, DICOM RASTER ENGINE (DRE).
  • Page 111: Error 20-209: Laser Imager Opened During Self-Test

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-209: LASER IMAGER Opened During Self-Test 8F2919 Page 111 of 326 Messages LOCAL PANEL DICOM SEVERITY Laser Imager Opened During Door or Drawer Opened PRINTER DOWN 0-Fatal Self-Test During Self-Test (Alarm sounds) Description This error occurs during MCS diagnostics when the IMAGER determines that a DOOR or a DRAWER was opened during running of the Self-Test.
  • Page 112 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 112 of 326 LATCH LATCH SENSOR Install a new LATCH POSITION SENSOR. See the ILLUSTRATED PARTS LIST, MAIN AY - Side Components. LATCH SPRING Install a new SPRING.
  • Page 113: Error 20-550: Preventive Maintenance Necessary

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-550: Preventive Maintenance Necessary 8F2919 Page 113 of 326 Messages LOCAL PANEL DICOM SEVERITY None Preventive Maintenance NORMAL 3-Notify (No Alarm) Required Description The PM message is written to the log when the sheet count exceeds the PM sheet count limit.
  • Page 114: Error 20-551: Processor Filter Maintenance Necessary

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-551: PROCESSOR FILTER Maintenance Necessary 8F2919 Page 114 of 326 Messages LOCAL PANEL DICOM SEVERITY None Processor Filter Maintenance NORMAL 3-Notify (No Alarm) Required Description The PM message is written to the log when the sheet count exceeds the sheet count limit for the Preventive Maintenance Module (PMM).
  • Page 115: Error 20-913: Dre Data Transfer Error Notification

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-913: DRE Data Transfer Error Notification 8F2919 Page 115 of 326 Messages LOCAL PANEL DICOM SEVERITY None Image transfer failure. Wrong NORMAL 3-Notify (No Alarm) number of bytes transferred in the allotted time frame. Description This error occurs when the wrong number of pixels is transferred between the DRE and the DPB in the allotted time frame.
  • Page 116: Error 20-915: Dre Data Transfer Fatal Error Detected

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-915: DRE Data Transfer Fatal Error Detected 8F2919 Page 116 of 326 Messages LOCAL PANEL DICOM SEVERITY Internal Image Data Transfer Image transfer failure after ELEC DOWN 0-Fatal Failed multiple attempts. Wrong (Alarm sounds) number of bytes delivered in the allotted time frame.
  • Page 117 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Install a new DRE. See the ADJUSTMENTS AND REPLACEMENTS, DICOM RASTER 8F2919 Page ENGINE (DRE). 117 of 326 Install a new OPTICS AY. See the ADJUSTMENTS AND REPLACEMENTS, OPTICS AY and EXPOSURE TRANSPORT...
  • Page 118: Error 20-917: Dre Data Transfer Timing Error Notification

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-917: DRE Data Transfer Timing Error Notification 8F2919 Page 118 of 326 Messages LOCAL PANEL DICOM SEVERITY None Image transfer failure. NORMAL 3-Notify (No Alarm) Transfer rate inadequate. Description This error occurs when the rate of image data transfer between the DRE and DPB is inadequate to feed the EXPOSURE TRANSPORT SUBSYSTEM.
  • Page 119: Error 20-918: Dre Data Transfer Too Slow. Fatal Timing Error Detected

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-918: DRE Data Transfer Too Slow. Fatal Timing Error Detected 8F2919 Page 119 of 326 Messages LOCAL PANEL DICOM SEVERITY Internal Image Data Transfer Image transfer failure after ELEC DOWN 0-Fatal Failed multiple attempts.
  • Page 120 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Install a new DRE. See the ADJUSTMENTS AND REPLACEMENTS, DICOM RASTER 8F2919 Page ENGINE (DRE). 120 of 326 Install a new OPTICS AY. See the ADJUSTMENTS AND REPLACEMENTS, OPTICS AY and EXPOSURE TRANSPORT...
  • Page 121: Error 20-919: Dre Image Render Not Complete

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 20-919: DRE Image Render Not Complete 8F2919 Page 121 of 326 Messages LOCAL PANEL DICOM SEVERITY Internal Image Data Render Image render did not ELEC DOWN 0-Fatal Failed complete (Alarm sounds) Description This error occurs when image data transfer between the DRE and the MCS does not complete within a the expected time.
  • Page 122 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Install a new OPTICS AY. See the ADJUSTMENTS AND REPLACEMENTS, OPTICS AY 8F2919 Page and EXPOSURE TRANSPORT 122 of 326...
  • Page 123: Error 21-116, 22-116, Or 23-116: Film Pickup Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-116, 22-116, or 23-116: Film Pickup Failure 8F2919 Page 123 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Jam in Area 1: xxxxx xxxxx Supply: Film Pickup FILM JAM 1-Alert Supply Failure. Film at Feed (S5) (Alarm sounds) sensor failed to detect film leading edge...
  • Page 124 DIAGNOSTICS Status Messages and Error Codes 30JUL07 FE Action 8F2919 Page Do the actions below in sequence. If an action provides a resolution, do not continue with the 124 of 326 other actions. De-energize then energize the LASER IMAGER. Do the following: Check: If an existing problem: FILM CARTRIDGE...
  • Page 125 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Check: If an existing problem: 8F2919 Page VACUUM PUMP Install a new PUMP. See the ADJUSTMENTS AND 125 of 326 REPLACEMENTS, VACUUM PUMP - PICKUP VACUUM VALVE Install a new VALVE. See the ADJUSTMENTS AND REPLACEMENTS, VACUUM RELIEF VALVE - PICKUP IDLER ROLLER TENSION...
  • Page 126: Error 21-117, 22-117, Or 23-117: Film Pickup Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-117, 22-117, or 23-117: Film Pickup Retry 8F2919 Page 126 of 326 Messages LOCAL PANEL DICOM SEVERITY None xxxxx Supply: Film Pickup NORMAL 3-Notify (No Alarm) Retry. Film at Feed Sensor (S5) failed to activate in specified time.
  • Page 127: Error 21-118, 22-118, Or 23-118: Pickup Did Not Go Home

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-118, 22-118, or 23-118: PICKUP Did Not Go Home 8F2919 Page 127 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Internal xxxxx Supply: Pickup Home FILM TRANS 1-Alert Hardware Failure Sensor (S1) failed to sense (Alarm sounds) pickup assembly.
  • Page 128 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 128 of 326 CONNECTORS and CABLES Seat all CONNECTORS. BLINDMATE CONNECTOR Adjust THE BLINDMATE. See the ADJUSTMENTS AND REPLACEMENTS, BLINDMATE BRACKET. PICKUP HOME SENSOR - Install a new SENSOR.
  • Page 129: Error 21-119, 22-119, Or 23-119: Film Contact Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-119, 22-119, or 23-119: Film Contact Failure 8F2919 Page 129 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Internal xxxxx Supply: Cups engaged FILM TRANS 1-Alert Hardware Failure sensor (S2) failed to confirm (Alarm sounds) that the pickup arm assembly lowered and contacted film...
  • Page 130 DIAGNOSTICS Status Messages and Error Codes 30JUL07 FE Action 8F2919 Page Do the actions below in sequence. If an action provides a resolution, do not continue with the 130 of 326 other actions. De-energize then energize the LASER IMAGER. Check the mechanical operation: •...
  • Page 131 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Check: If an existing problem: 8F2919 Page WIRING HARNESS Install a new HARNESS. See the ILLUSTRATED PARTS 131 of 326 LIST, PICKUP AY - HARNESS and TUBING. Install a new FSB. See the ADJUSTMENTS AND REPLACEMENTS, FILM SUPPLY BOARD (FSB) AY - PICKUP...
  • Page 132: Error 21-120, 22-120, Or 23-120: Cups Engaged Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-120, 22-120, or 23-120: CUPS Engaged Failure 8F2919 Page 132 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: internal xxxxx Supply: Cups engaged FILM TRANS 1-Alert Hardware Failure sensor (S2) failed to confirm (Alarm sounds) proper completion of the vacuum cup arm movement...
  • Page 133 DIAGNOSTICS Status Messages and Error Codes 30JUL07 FE Action 8F2919 Page Do the actions below in sequence. If an action provides a resolution, do not continue with the 133 of 326 other actions. De-energize then energize the LASER IMAGER. Do the following: Check: If an existing problem: If HEEL LOCK is engaged...
  • Page 134: Error 21-125, 22-125, Or 23-125: Feed Roller Nip Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-125, 22-125, or 23-125: FEED ROLLER Nip Failure 8F2919 Page 134 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Internal xxxxx Supply: Feed Roller FILM TRANS 1-Alert Hardware Failure Sensor (S4) failed to detect (Alarm sounds) feed roller open/closed position.
  • Page 135 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 135 of 326 FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. FEED ROLLER POSITION Install a new SENSOR. See the ADJUSTMENTS AND SENSOR - use the SERVICE REPLACEMENTS, SENSOR, FEED ROLLER...
  • Page 136: Error 21-126, 22-126, Or 23-126: Jam At Feeder

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-126, 22-126, or 23-126: Jam at Feeder 8F2919 Page 136 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Jam in Area 1: xxxxx xxxxx Supply: Film Jam - FILM JAM 1-Alert Supply Film at Feed Sensor (S5) (Alarm sounds) failed to activate in specified...
  • Page 137 DIAGNOSTICS Status Messages and Error Codes 30JUL07 FE Action 8F2919 Page Do the actions below in sequence. If an action provides a resolution, do not continue with the 137 of 326 other actions. De-energize then energize the LASER IMAGER. Do the following: Check: If an existing problem: FILM CARTRIDGE...
  • Page 138: Error 21-139, 22-139, Or 23-139: Unable To Read Or Write To Rf Tag

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-139, 22-139, or 23-139: Unable to Read or Write to RF TAG 8F2919 Page 138 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Unable to xxxxx Supply: RF Tag Reader BAD SUPPLY 1-Alert Identify Film Cartridge failed to Read or Write RF...
  • Page 139 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 139 of 326 FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. BLINDMATE CONNECTOR Adjust THE BLINDMATE. See the ADJUSTMENTS AND REPLACEMENTS, BLINDMATE BRACKET.
  • Page 140: Error 21-145, 22-145, Or 23-145: Unsupported Media Type

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-145, 22-145, or 23-145: Unsupported Media Type 8F2919 Page 140 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Unsupported xxxxx Supply: Media Type BAD SUPPLY 1-Alert Media Type located in cartridge is not (Alarm sounds) supported Description...
  • Page 141 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 141 of 326 FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. BLINDMATE CONNECTOR Adjust THE BLINDMATE. See the ADJUSTMENTS AND REPLACEMENTS, BLINDMATE BRACKET.
  • Page 142: Error 21-146, 22-146, Or 23-146: Unsupported Media Size

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-146, 22-146, or 23-146: Unsupported Media Size 8F2919 Page 142 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Unsupported xxxxx Supply: Media Size BAD SUPPLY 1-Alert Media Size located in cartridge is not (Alarm sounds) supported.
  • Page 143 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 143 of 326 FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. BLINDMATE CONNECTOR Adjust THE BLINDMATE. See the ADJUSTMENTS AND REPLACEMENTS, BLINDMATE BRACKET.
  • Page 144: Error 21-154, 22-154, Or 23-154: Supply Controller Nvram Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-154, 22-154, or 23-154: Supply CONTROLLER NVRAM Failure 8F2919 Page 144 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Internal xxxxx Supply: Microcontroller BAD SUPPLY 1-Alert Communications Failure failed to read/write to NVRAM (Alarm sounds) Description Note...
  • Page 145 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 145 of 326 FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. BLINDMATE CONNECTOR Adjust THE BLINDMATE. See the ADJUSTMENTS AND REPLACEMENTS, BLINDMATE BRACKET.
  • Page 146: Error 21-175, 22-175, Or 23-175: Rollback Did Not Engage Cartridge

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-175, 22-175, or 23-175: ROLLBACK Did Not Engage CARTRIDGE 8F2919 Page 146 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Hardware or xxxxx Supply: Rollback Failed BAD SUPPLY 1-Alert Cartridge Failure to Engage Cartridge (Alarm sounds) Description...
  • Page 147 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 147 of 326 FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. BLINDMATE CONNECTOR Adjust THE BLINDMATE. See the ADJUSTMENTS AND REPLACEMENTS, BLINDMATE BRACKET.
  • Page 148: Error 21-176, 22-176, Or 23-176: Film Cartridge Did Not Open Fully

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-176, 22-176, or 23-176: FILM CARTRIDGE Did Not Open Fully 8F2919 Page 148 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Film Cartridge xxxxx Supply: Cartridge BAD SUPPLY 1-Alert Failed to Open Open Sensor (S7) failed to (Alarm sounds) activate in specified time...
  • Page 149 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 149 of 326 For obstructions Remove obstructions. For film jam Remove films. FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. BLINDMATE CONNECTOR Adjust THE BLINDMATE.
  • Page 150: Error 21-177, 22-177, Or 23-177: Film Cartridge Did Not Close

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-177, 22-177, or 23-177: FILM CARTRIDGE Did Not Close 8F2919 Page 150 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Film Cartridge xxxxx Supply: Rollback Home BAD SUPPLY 1-Alert Failed to Close Sensor (S6) failed to activate (Alarm sounds) in specified time...
  • Page 151 DIAGNOSTICS Status Messages and Error Codes 30JUL07 FE Action 8F2919 Page Do the actions below in sequence. If an action provides a resolution, do not continue with the 151 of 326 other actions. De-energize then energize the LASER IMAGER. Do the following: Check: If an existing problem: For obstructions...
  • Page 152 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Check: If an existing problem: 8F2919 Page CHAIN for obstructions or Remove obstruction. 152 of 326 damage Install a new CHAIN. See the ADJUSTMENTS AND REPLACEMENTS, CHAIN - ROLLBACK Install a new FSB. See the ADJUSTMENTS AND REPLACEMENTS, FILM SUPPLY BOARD (FSB) AY - PICKUP...
  • Page 153: Error 21-178, 22-178, Or 23-178: Rollback Did Not Leave Home

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-178, 22-178, or 23-178: ROLLBACK Did Not Leave Home 8F2919 Page 153 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Film Cartridge xxxxx Supply: Rollback Home BAD SUPPLY 1-Alert Failed to Open Sensor (S6) failed to (Alarm sounds) deactivate in specified time...
  • Page 154 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 154 of 326 For obstructions Remove obstructions. For film jam Remove films. FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. BLINDMATE CONNECTOR Adjust THE BLINDMATE.
  • Page 155 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Install a new ROLLBACK AY. See the ADJUSTMENTS AND REPLACEMENTS, 8F2919 Page ROLLBACK 155 of 326...
  • Page 156: Error 21-179, 22-179, Or 23-179: Rollback Did Not Move From Open

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-179, 22-179, or 23-179: ROLLBACK Did Not Move From Open 8F2919 Page 156 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Film Cartridge xxxxx Supply: Cartridge BAD SUPPLY 1-Alert Failed to Close Open Sensor (S7) failed to (Alarm sounds) deactivate in specified time...
  • Page 157 DIAGNOSTICS Status Messages and Error Codes 30JUL07 FE Action 8F2919 Page Do the actions below in sequence. If an action provides a resolution, do not continue with the 157 of 326 other actions. De-energize then energize the LASER IMAGER. Do the following: Check: If an existing problem: For obstructions...
  • Page 158 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Check: If an existing problem: 8F2919 Page CHAIN for obstructions or Remove obstruction. 158 of 326 damage Install a new CHAIN. See the ADJUSTMENTS AND REPLACEMENTS, CHAIN - ROLLBACK Install a new FSB. See the ADJUSTMENTS AND REPLACEMENTS, FILM SUPPLY BOARD (FSB) AY - PICKUP...
  • Page 159: Error 21-320, 22-320, Or 23-320: Jam Entering The Accumulator

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-320, 22-320, or 23-320: Jam Entering the ACCUMULATOR 8F2919 Page 159 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Jam in Area 2 or Area xxxxx Supply: Film Jam FILM JAM 1-Alert 1:xxxxx Supply Entering Accumulator: (Alarm sounds)
  • Page 160 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Close all DOORS and the FILM DRAWER. 8F2919 Page If the error occurs again, call Service. 160 of 326 FE Action Do the actions below in sequence. If an action provides a resolution, do not continue with the other actions.
  • Page 161: Error 21-624, 22-624, Or 23-624: Bad Densitometer Data

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-624, 22-624, or 23-624: Bad DENSITOMETER Data 8F2919 Page 161 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Film xxxxx Supply: Calibration CALIBRATION 1-Alert Calibration Failure (Alarm failed sounds) Description Note UPPER DRAWER (U) = 21.
  • Page 162 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 162 of 326 FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. BELTS on TURNAROUND Install a new BELT. ROLLERS on TURNAROUND Install a new TURNAROUND AY.
  • Page 163: Error 21-631, 22-631, Or 23-631: Dmin Not Met

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-631, 22-631, or 23-631: Dmin Not Met 8F2919 Page 163 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Film xxxxx Supply: Calibration CALIBRATION 1-Alert Calibration Failure - Dmin failed - Dmin Not Met Outside Target (Alarm sounds) Description...
  • Page 164 DIAGNOSTICS Status Messages and Error Codes 30JUL07 De-energize then energize the LASER IMAGER. 8F2919 Page Do the following: 164 of 326 Check: If an existing problem: FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. ROLLERS on TURNAROUND Install a new TURNAROUND AY. See the ADJUSTMENTS AND REPLACEMENTS, TURNAROUND AY - IMAGER with SORTER.
  • Page 165: Error 21-632, 22-632, Or 23-632: Dmax Not Met

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-632, 22-632, or 23-632: Dmax Not Met 8F2919 Page 165 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Film xxxxx Supply: Calibration CALIBRATION 1-Alert Calibration Failure - Dmax failed - Dmax Not Met Outside Target (Alarm sounds) Description...
  • Page 166 DIAGNOSTICS Status Messages and Error Codes 30JUL07 De-energize then energize the LASER IMAGER. 8F2919 Page Do the following: 166 of 326 Check: If an existing problem: FILM CARTRIDGE Install a new FILM CARTRIDGE. CONNECTORS and CABLES Seat all CONNECTORS. ROLLERS on TURNAROUND Install a new TURNAROUND AY. See the ADJUSTMENTS AND REPLACEMENTS, TURNAROUND AY - IMAGER with SORTER.
  • Page 167: Error 21-922, 22-922, Or 23-922: Rf Tag Diagnostics Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-922, 22-922, or 23-922: RF TAG Diagnostics Failure 8F2919 Page 167 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Internal xxxxx Supply: RF Tag reader BAD SUPPLY 1-Alert Diagnostics Failure - RF Tag failed self diagnostics (Alarm sounds) Description...
  • Page 168 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 168 of 326 FILM CARTRIDGE Install a new FILM CARTRIDGE. RF TAG ANTENNA CABLE Seat the CONNECTOR. Install a new HARNESS. See the ADJUSTMENTS AND REPLACEMENTS, HARNESS - ROLLBACK RF ANTENNA BOARD...
  • Page 169: Error 21-931, 22-931, Or 23-931: Supply Controller Communications Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-931, 22-931, or 23-931: SUPPLY CONTROLLER Communications 8F2919 Page Failure 169 of 326 Messages LOCAL PANEL DICOM SEVERITY xxxxx Supply: Internal MCS failed to communicate BAD SUPPLY 1-Alert Communications Failure with xxxxx Supply after (Alarm sounds) multiple attempts Description...
  • Page 170 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 170 of 326 CONNECTORS and CABLES Seat all CONNECTORS. to/from the PDB BLINDMATE CONNECTOR Adjust THE BLINDMATE. See the ADJUSTMENTS AND REPLACEMENTS, BLINDMATE BRACKET.
  • Page 171: Error 21-932, 22-932, Or 23-932: Supply Controller Communications Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 21-932, 22-932, or 23-932: SUPPLY CONTROLLER Communications 8F2919 Page Retry 171 of 326 Messages LOCAL PANEL DICOM SEVERITY None MCS failed to communicate NORMAL 3-Notify (No alarm) with xxxxx Supply Description Note UPPER DRAWER (U) = 21.
  • Page 172: Error 26-321: Jam Exiting Accumulator

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 26-321: Jam Exiting ACCUMULATOR 8F2919 Page 172 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Jam in Area 2 Film Jam Exiting FILM JAM 1-Alert (Alarm sounds) Accumulator: Accumulator Sensor S9 failed to detect film trailing edge Description This error occurs if the trailing edge is not detected at the FILM REGISTRATION...
  • Page 173 DIAGNOSTICS Status Messages and Error Codes 30JUL07 FE Action 8F2919 Page Do the actions below in sequence. If an action provides a resolution, do not continue with the 173 of 326 other actions. De-energize then energize the LASER IMAGER. Do the following: Check: If an existing problem: For film jam in area 2...
  • Page 174: Error 26-325: Jam Entering Exposure Transport

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 26-325: Jam Entering EXPOSURE TRANSPORT 8F2919 Page 174 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Jam in Area 2 Film Jam Entering Exposure FILM JAM 1-Alert (Alarm sounds) Transport: Exposure Entrance Sensor S12 failed to detect film leading edge Description...
  • Page 175 DIAGNOSTICS Status Messages and Error Codes 30JUL07 FE Action 8F2919 Page Do the actions below in sequence. If an action provides a resolution, do not continue with the 175 of 326 other actions. De-energize then energize the LASER IMAGER. Do the following: Check: If an existing problem: For film jam in area 2...
  • Page 176: Error 26-326: Jam Exiting Exposure Transport

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 26-326: Jam Exiting EXPOSURE TRANSPORT 8F2919 Page 176 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Jam in Area 2 or 3 Film Jam Exiting Exposure FILM JAM 1-Alert (Alarm sounds) Transport: Exposure Entrance Sensor S12 failed to detect film trailing edge Description...
  • Page 177 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Remove any films in area 3. 8F2919 Page 177 of 326 Note Use the FILM FEED KNOB in area 3 to move the film into area 3. “Relatch” and close the PROCESSOR ASSEMBLY. Close all DOORS.
  • Page 178: Error 26-543: Jam Before The Densitometer

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 26-543: Jam Before the DENSITOMETER 8F2919 Page 178 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Jam in Area 3 or 4 Film Jam Prior to FILM JAM 1-Alert (Alarm sounds) Densitometer: Film at Densitometer sensor (S18) failed to detect film leading edge...
  • Page 179 DIAGNOSTICS Status Messages and Error Codes 30JUL07 • Use the FILM FEED KNOB in area 3 to advance the film. 8F2919 Page • “Unlatch” and open the PROCESSOR ASSEMBLY 179 of 326 • Remove any films in area 3. • “Relatch”...
  • Page 180: Error 26-544: Jam At Densitometer

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 26-544: Jam at DENSITOMETER 8F2919 Page 180 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Jam in Area 5 Film Jam at Densitometer: FILM JAM 1-Alert (Alarm sounds) Film at Densitometer sensor (S18) failed to detect film trailing edge Description...
  • Page 181 DIAGNOSTICS Status Messages and Error Codes 30JUL07 FE Action 8F2919 Page Do the actions below in sequence. If an action provides a resolution, do not continue with the 181 of 326 other actions. De-energize then energize the LASER IMAGER. Do the following: Check: If an existing problem: For film jam in area 5...
  • Page 182: Error 27-121: Attenuator Motor Home Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-121: ATTENUATOR MOTOR Home Error 8F2919 Page 182 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Internal Hardware Attenuator Home Sensor ELEC DOWN 0-Fatal Failure (S24) failed to detect (Alarm sounds) attenuator motor movement Description This error occurs if the OPTICS ATTENUATOR MOTOR (M22) does not move to or from the home position when requested, as indicated by the OPTICS ATTENUATOR HOME SENSOR...
  • Page 183 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 183 of 326 CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE • • • SOP BOARD OPTICS ATTENUATOR Install a new OPTICS AY. See the ADJUSTMENTS AND HOME SENSOR - use the REPLACEMENTS, OPTICS AY and EXPOSURE...
  • Page 184: Error 27-123: Polygon Motor Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-123: POLYGON MOTOR Error 8F2919 Page 184 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Internal Hardware Polygon fails to rotate at ELEC DOWN 0-Fatal Failure desired speed (Alarm sounds) Description This error occurs when the POLYGON speed measured by the OPTICS MODULE MICROCONTROLLER is out of range.
  • Page 185 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 185 of 326 For open DOORS or Close all DOORS and DRAWERS. DRAWERS CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE •...
  • Page 186: Error 27-154: Optics Nvram Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-154: OPTICS NVRAM Failure 8F2919 Page 186 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Internal Optics (BPM) microcontroller ELEC DOWN 0-Fatal Communications Failure failed to read/write to NVRAM (Alarm sounds) Description This error occurs when the MICROCONTROLLER detects an error when attempting to read or write from NVRAM.
  • Page 187 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 187 of 326 CONNECTORS on the: Seat the CONNECTORS. • • • Install a new PDB. See the ADJUSTMENTS AND REPLACEMENTS, POWER DISTRIBUTION BOARD (PDB).
  • Page 188: Error 27-601: Invalid Power Monitor Offset

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-601: Invalid POWER MONITOR Offset 8F2919 Page 188 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Calibration Failed Optics Calibration Failed. ELEC DOWN 0-Fatal (Alarm sounds) Power Monitor Offset outside valid range Description This error occurs when the POWER MONITOR offset measured by the OPTICS calibration and diagnostics is outside the range.
  • Page 189 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 189 of 326 For open DOORS or Close all DOORS and DRAWERS. DRAWERS CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE •...
  • Page 190: Error 27-602: Invalid Power Monitor Range With Attenuator Open

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-602: Invalid POWER MONITOR Range with ATTENUATOR Open 8F2919 Page 190 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Calibration Failed Optics Calibration Failed. ELEC DOWN 0-Fatal (Alarm sounds) Power Monitor value with Attenuator Open outside valid range Description...
  • Page 191 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 191 of 326 For open DOORS or Close all DOORS and DRAWERS. DRAWERS CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE •...
  • Page 192: Error 27-603: Invalid Attenuator Optical Density Range

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-603: Invalid ATTENUATOR Optical Density Range 8F2919 Page 192 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Calibration Failed Optics Calibration Failed. ELEC DOWN 0-Fatal (Alarm sounds) Attenuator Optical Density value outside valid range Description This error occurs when the ATTENUATOR optical density range measured by the OPTICS calibration and diagnostics is outside the range.
  • Page 193 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 193 of 326 For open DOORS or Close all DOORS and DRAWERS. DRAWERS CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE •...
  • Page 194: Error 27-604: Unable To Adjust Laser Dynamic Range

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-604: Unable to Adjust LASER Dynamic Range 8F2919 Page 194 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Calibration Failed Optics calibration failed to ELEC DOWN 0-Fatal (Alarm sounds) Adjust Laser Dynamic Range within limits Description This error occurs when the LASER dynamic range measured by the OPTICS calibration and...
  • Page 195 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 195 of 326 For open DOORS or Close all DOORS and DRAWERS. DRAWERS CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE •...
  • Page 196: Error 27-605: Power Monitor Saturation

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-605: POWER MONITOR Saturation 8F2919 Page 196 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Calibration Failed Optics Calibration detected ELEC DOWN 0-Fatal (Alarm sounds) Power Monitor Saturation Description This error occurs when the POWER MONITOR is saturated with the LASER set at maximum power.
  • Page 197 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 197 of 326 For open DOORS or Close all DOORS and DRAWERS. DRAWERS CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE •...
  • Page 198: Error 27-606: Attenuator Calibration Failed (Non-Monotonic)

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-606: ATTENUATOR Calibration Failed (Non-Monotonic) 8F2919 Page 198 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Calibration Failed Optics Calibration Failed. ELEC DOWN 0-Fatal (Alarm sounds) Power Monitor readings not monotonic Description This error occurs when the BEAM POWER measurements taken during ATTENUATOR calibration are non-monotonic.
  • Page 199 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 199 of 326 For open DOORS or Close all DOORS and DRAWERS. DRAWERS CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE •...
  • Page 200: Error 27-607: Optics Exceeded Maximum Laser Power

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-607: OPTICS Exceeded Maximum LASER Power 8F2919 Page 200 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Calibration Failed Optics Calibration detected ELEC DOWN 0-Fatal (Alarm sounds) Optics Maximum Power outside valid range Description This error occurs when the maximum LASER power measured by the OPTICS calibration and diagnostics is outside the range.
  • Page 201 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 201 of 326 For open DOORS or Close all DOORS and DRAWERS. DRAWERS CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE •...
  • Page 202: Error 27-608: Invalid Laser Dynamic Range

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-608: Invalid LASER Dynamic Range 8F2919 Page 202 of 326 Messages LOCAL PANEL DICOM SEVERITY None Optics Calibration detected NORMAL 3-Notify (No alarm) laser dynamic range approaching predefined limits Description This occurs when the laser dynamic range is adjusted and the value is within 10% of the adjustment range.
  • Page 203: Error 27-609: Optics Approaching Maximum Laser Power

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-609: OPTICS Approaching Maximum LASER Power 8F2919 Page 203 of 326 Messages LOCAL PANEL DICOM SEVERITY None Optics Calibration detected NORMAL 3-Notify (No alarm) Optics Maximum Power approaching limits Description This occurs if the maximum LASER power measured by OPTICS calibration and diagnostics is approaching the limit.
  • Page 204: Error 27-611: Optics Laser Gain Adjustment Failed

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-611: OPTICS LASER Gain Adjustment Failed 8F2919 Page 204 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Internal Hardware Optics calibration laser power ELEC DOWN 0-Fatal Failure gain adjustment failed (Alarm sounds) Description The laser power gain is measured using a voltage on the LASER DRIVER.
  • Page 205 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 205 of 326 For open DOORS or Close all DOORS and DRAWERS. DRAWERS CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE •...
  • Page 206: Error 27-646: Start Of Page Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-646: Start of Page Error 8F2919 Page 206 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Internal Hardware Start of Page Sensor (S13) ELEC DOWN 0-Fatal Failure failed to detect edge of film (Alarm sounds) Description This error occurs when the OPTICS MODULE does not detect a start of page signal as...
  • Page 207 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 207 of 326 CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE • • • SOP BOARD SOP BOARD - use the Install a new SENSOR.
  • Page 208: Error 27-647: Start Of Page Sensor Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-647: START OF PAGE SENSOR Failure 8F2919 Page 208 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Internal Hardware Start of Page Sensor (S13) ELEC DOWN 0-Fatal Failure failed diagnostics (Alarm sounds) Description This error occurs when the START OF PAGE SENSOR (S13) does not detect the normal LASER on or off state during diagnostics.
  • Page 209 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 209 of 326 CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE • • • SOP BOARD SOP BOARD - use the Install a new SENSOR.
  • Page 210: Error 27-648: Laser Scan Correction Dac Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-648: LASER Scan Correction DAC Failure 8F2919 Page 210 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Internal Hardware Laser Scan Correction DAC ELEC DOWN 0-Fatal Failure failed diagnostics (Alarm sounds) Description This error occurs during LASER scan correction DAC diagnostics.
  • Page 211 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 211 of 326 CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE • • • SOP BOARD Install a new ETB. See the ADJUSTMENTS AND REPLACEMENTS, EXPOSURE TRANSPORT BOARD (ETB) - EXPOSURE TRANSPORT...
  • Page 212: Error 27-650: Exposure Timeout Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-650: Exposure Timeout Error 8F2919 Page 212 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Internal Hardware Image Exposure time out ELEC DOWN 0-Fatal Failure failure -- multiple successive (Alarm sounds) timeouts Description This error occurs when the image exposure does not complete in the necessary time for 3 successive films.
  • Page 213 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 213 of 326 CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE • • • SOP BOARD Install a new ETB. See the ADJUSTMENTS AND REPLACEMENTS, EXPOSURE TRANSPORT BOARD (ETB) - EXPOSURE TRANSPORT...
  • Page 214: Error 27-651: Exposure Timeout Notification

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-651: Exposure Timeout Notification 8F2919 Page 214 of 326 Messages LOCAL PANEL DICOM SEVERITY None Image Exposure timed out, NORMAL 3-Notify (No alarm) job will be reprinted Description This error occurs when the image exposure does not complete in the necessary time. The faulty image will be printed.
  • Page 215: Error 27-931: Optics Communications Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-931: OPTICS Communications Failure 8F2919 Page 215 of 326 Messages LOCAL PANEL DICOM SEVERITY Optics: Internal MCS failed to communicate ELEC DOWN 0-Fatal Communications Failure with Optics (BPM) Module (Alarm sounds) after multiple attempts Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response to 5 attempts.
  • Page 216 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 216 of 326 CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE • • • SOP BOARD Install a new ETB. See the ADJUSTMENTS AND REPLACEMENTS, EXPOSURE TRANSPORT BOARD (ETB) - EXPOSURE TRANSPORT...
  • Page 217: Error 27-932: Optics Communications Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 27-932: OPTICS Communications Retry 8F2919 Page 217 of 326 Messages LOCAL PANEL DICOM SEVERITY None MCS failed to communicate NORMAL 3-Notify (No alarm) with Optics (BPM) Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response.
  • Page 218: Error 28-154: Processor Nvram Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 28-154: PROCESSOR NVRAM Failure 8F2919 Page 218 of 326 Messages LOCAL PANEL DICOM SEVERITY Processor: Internal Processor microcontroller ELEC DOWN 0-Fatal Communications Failure failed to read/write to NVRAM (A beep occurs) Description This error occurs when the MICROCONTROLLER detect an error when attempting to read or write from NVRAM.
  • Page 219 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 219 of 326 CONNECTORS on the: Seat the CONNECTORS. • DRE USB CABLE • • Install a new PCB. See the ADJUSTMENTS AND REPLACEMENTS, PROCESSOR CONTROL BOARD (PCB) - THERMAL PROCESSOR...
  • Page 220: Error 28-155: Bad Processor Nvram Parameters

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 28-155: Bad PROCESSOR NVRAM Parameters 8F2919 Page 220 of 326 Messages LOCAL PANEL DICOM SEVERITY Processor: Internal Invalid Processor parameters PROC DOWN 0-Fatal Communications Failure read from NVRAM (Alarm sounds) Description This error occurs during start up when the setpoint or offset parameters read from NVRAM are outside the allowable values.
  • Page 221 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 221 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new PCB. See the ADJUSTMENTS AND REPLACEMENTS, PROCESSOR CONTROL BOARD (PCB) - THERMAL PROCESSOR Install a new PDB.
  • Page 222: Error 28-501: Processor Rtd Short Circuit

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 28-501: PROCESSOR RTD Short Circuit 8F2919 Page 222 of 326 Messages LOCAL PANEL DICOM SEVERITY Processor: Internal Hardware Processor calculates PROC DOWN 0-Fatal Failure temperature near zero. (Alarm sounds) Probably short in RTD circuit Description This error occurs when the temperature calculated by the PROCESSOR MICROPROCESSOR is almost zero.
  • Page 223 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 223 of 326 CONNECTORS on the: Seat the CONNECTORS. • • CIRCUIT BREAKER on the Engage the CIRCUIT BREAKER. POWER SUPPLY RTD settings - use the Install a new PROCESSOR AY.
  • Page 224: Error 28-509: Processor Did Not Complete Warm Up

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 28-509: PROCESSOR Did Not Complete Warm Up 8F2919 Page 224 of 326 Messages LOCAL PANEL DICOM SEVERITY Processor Warm-up Failure Processor failed to warm up PROC DOWN 0-Fatal (Alarm sounds) in the specified time Description This error occurs if any PROCESSOR RTD indicates a temperature lower than the setpoint temperature when the allotted warm up time has elapsed.
  • Page 225 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 225 of 326 CONNECTORS on the: Seat the CONNECTORS. • • CIRCUIT BREAKER on the Engage the CIRCUIT BREAKER. POWER SUPPLY RTD settings - use the Install a new PROCESSOR AY.
  • Page 226: Error 28-551: Processor Heater Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 28-551: PROCESSOR HEATER Failure 8F2919 Page 226 of 326 Messages LOCAL PANEL DICOM SEVERITY Processor Heater Failure Processor failed to warm up PROC DOWN 0-Fatal (Alarm sounds) when heater turned on Description This error occurs if any PROCESSOR RTD indicates that the temperature is not increasing at the expected rate.
  • Page 227 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 227 of 326 CONNECTORS on the: Seat the CONNECTORS. • • CIRCUIT BREAKER on the Engage the CIRCUIT BREAKER. POWER SUPPLY RTD settings - use the Install a new PROCESSOR AY.
  • Page 228: Error 28-554: Temperature Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 28-554: Temperature Error 8F2919 Page 228 of 326 Messages LOCAL PANEL DICOM SEVERITY Processor Over Temperature Processor Over Temperature PROC DOWN 0-Fatal (Alarm sounds) Description This error occurs if any PROCESSOR RTD indicates that the temperature is higher than the limit.
  • Page 229 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 229 of 326 Ambient room temperature “Notify” customer that ambient temperature is not correct the operation of the LASER IMAGER. CONNECTORS on the: Seat the CONNECTORS.
  • Page 230: Error 28-931: Processor Communications Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 28-931: PROCESSOR Communications Failure 8F2919 Page 230 of 326 Messages LOCAL PANEL DICOM SEVERITY Processor: Internal MCS failed to communicate ELEC DOWN 0-Fatal Communications Failure with the processor module (Alarm sounds) after multiple attempts Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response to 5 attempts.
  • Page 231 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 231 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new PCB. See the ADJUSTMENTS AND REPLACEMENTS, PROCESSOR CONTROL BOARD (PCB) - THERMAL PROCESSOR Install a new PDB.
  • Page 232: Error 28-932: Processor Communications Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 28-932: PROCESSOR Communications Retry 8F2919 Page 232 of 326 Messages LOCAL PANEL DICOM SEVERITY None MCS failed to communicate NORMAL 3-Notify (No alarm) with the Processor Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response.
  • Page 233: Error 29-154: Densitometer Nvram Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 29-154: DENSITOMETER NVRAM Failure 8F2919 Page 233 of 326 Messages LOCAL PANEL DICOM SEVERITY Densitometer: Internal Densitometer microcontroller ELEC DOWN 0-Fatal Communications Failure failed to read/write to NVRAM (Alarm sounds) Description This error occurs when the MICROCONTROLLER detects an error when attempting to read or write from NVRAM.
  • Page 234 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 234 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new PCB. See the ADJUSTMENTS AND REPLACEMENTS, PROCESSOR CONTROL BOARD (PCB) - THERMAL PROCESSOR Install a new PDB.
  • Page 235: Error 29-924: Densitometer Failed Offset Reading Diagnostics

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 29-924: DENSITOMETER Failed Offset Reading Diagnostics 8F2919 Page 235 of 326 Messages LOCAL PANEL DICOM SEVERITY Densitometer: Internal Densitometer offset reading ELEC DOWN 0-Fatal Diagnostics Failure out of range with light source (Alarm sounds) Description During a calibration print and during DENSITOMETER diagnostics, a reading from the...
  • Page 236 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 236 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new PCB. See the ADJUSTMENTS AND REPLACEMENTS, PROCESSOR CONTROL BOARD (PCB) - THERMAL PROCESSOR Install a new PDB.
  • Page 237: Error 29-925: Densitometer Reference Reading Failed Diagnostics

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 29-925: DENSITOMETER Reference Reading Failed Diagnostics 8F2919 Page 237 of 326 Messages LOCAL PANEL DICOM SEVERITY Densitometer: Internal Densitometer reference ELEC DOWN 0-Fatal Diagnostics Failure reading out of range with light (Alarm sounds) source on Description During a calibration print and during DENSITOMETER diagnostics, a reading from the...
  • Page 238 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 238 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new PCB. See the ADJUSTMENTS AND REPLACEMENTS, PROCESSOR CONTROL BOARD (PCB) - THERMAL PROCESSOR Install a new PDB.
  • Page 239: Error 29-926: Densitometer Fiducial Failed Offset Diagnostics

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 29-926: DENSITOMETER Fiducial Failed Offset Diagnostics 8F2919 Page 239 of 326 Messages LOCAL PANEL DICOM SEVERITY Densitometer: Internal Densitometer fiducial sensor ELEC DOWN 0-Fatal Diagnostics Failure offset reading out of range (Alarm sounds) with light source off Description During a calibration print and during DENSITOMETER diagnostics, a reading from the...
  • Page 240 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 240 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new PCB. See the ADJUSTMENTS AND REPLACEMENTS, PROCESSOR CONTROL BOARD (PCB) - THERMAL PROCESSOR Install a new PDB.
  • Page 241: Error 29-927: Densitometer Fiducial Failed Reference Diagnostics

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 29-927: DENSITOMETER Fiducial Failed Reference Diagnostics 8F2919 Page 241 of 326 Messages LOCAL PANEL DICOM SEVERITY Densitometer: Internal Densitometer fiducial sensor ELEC DOWN 0-Fatal Diagnostic Failure reference reading out of (Alarm sounds) range with light source on Description During a calibration print and during DENSITOMETER diagnostics, a reading from the...
  • Page 242 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 242 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new PCB. See the ADJUSTMENTS AND REPLACEMENTS, PROCESSOR CONTROL BOARD (PCB) - THERMAL PROCESSOR Install a new PDB.
  • Page 243: Error 29-931: Densitometer Communications Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 29-931: DENSITOMETER Communications Failure 8F2919 Page 243 of 326 Messages LOCAL PANEL DICOM SEVERITY Densitometer: Internal MCS failed to communicate ELEC DOWN 0-Fatal Communications Failure with Densitometer after (Alarm sounds) multiple attempts Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response to 5 attempts.
  • Page 244 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 244 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new PCB. See the ADJUSTMENTS AND REPLACEMENTS, PROCESSOR CONTROL BOARD (PCB) - THERMAL PROCESSOR Install a new PDB.
  • Page 245: Error 29-932: Densitometer Communications Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 29-932: DENSITOMETER Communications Retry 8F2919 Page 245 of 326 Messages LOCAL PANEL DICOM SEVERITY None MCS failed to communicate NORMAL 3-Notify (No alarm) with Densitometer Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response.
  • Page 246: Error 29-945: Densitometer Reference Reading Diagnostics Warning

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 29-945: DENSITOMETER Reference Reading Diagnostics Warning 8F2919 Page 246 of 326 Messages LOCAL PANEL DICOM SEVERITY None Densitometer reference NORMAL 3-Notify (No alarm) reading outside of optimal range with light source on Description During a calibration print and during DENSITOMETER diagnostics, a reading from the DENSITY SENSOR (S16) is taken with the light source turned on.
  • Page 247: Error 29-947: Densitometer Fiducial Reference Diagnostics Warning

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 29-947: DENSITOMETER Fiducial Reference Diagnostics Warning 8F2919 Page 247 of 326 Messages LOCAL PANEL DICOM SEVERITY None Densitometer fiducial sensor NORMAL 3-Notify (No alarm) reference reading outside of optimal range with light source on Description During a calibration print and during DENSITOMETER diagnostics, a reading from the...
  • Page 248: Error 30-154: Sorter Nvram Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 30-154: SORTER NVRAM Failure 8F2919 Page 248 of 326 Messages LOCAL PANEL DICOM SEVERITY Sorter: Internal Sorter microcontroller failed ELEC DOWN 1-Alert Communications Failure to read/write to NVRAM (Alarm sounds) Description This error occurs when the MICROCONTROLLER detects an error when attempting to read or write from NVRAM.
  • Page 249 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 249 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new SCB. Install a new PDB. See the ADJUSTMENTS AND REPLACEMENTS, POWER DISTRIBUTION BOARD (PDB).
  • Page 250: Error 30-931: Sorter Communications Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 30-931: SORTER Communications Failure 8F2919 Page 250 of 326 Messages LOCAL PANEL DICOM SEVERITY Sorter: Internal MCS failed to communicate ELEC DOWN 1-Alert Communications Failure with Sorter after multiple (Alarm sounds) attempts Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response to 5 attempts.
  • Page 251 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 251 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new SCB. Install a new PDB. See the ADJUSTMENTS AND REPLACEMENTS, POWER DISTRIBUTION BOARD (PDB).
  • Page 252: Error 30-932: Sorter Communications Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 30-932: SORTER Communications Retry 8F2919 Page 252 of 326 Messages LOCAL PANEL DICOM SEVERITY None MCS failed to communicate NORMAL 3-Notify (No Alarm) with Sorter Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response.
  • Page 253: Error 30-935: Sorter Diverter #1 Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 30-935: SORTER DIVERTER #1 Failure 8F2919 Page 253 of 326 Messages LOCAL PANEL DICOM SEVERITY Sorter: Internal Hardware Sorter Diverter #1 failed to ELEC DOWN 1-Alert Failure operate (Alarm sounds) Description This error occurs when the SORTER DIVERTER ARM does not move, as determined by the DIVERTER BIN 1 POSITION SENSOR (S20).
  • Page 254 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 254 of 326 CONNECTORS on the: Seat the CONNECTORS. • DIVERTER MOTOR • • DIVERTER MOTOR Install a new MOTOR. See the ADJUSTMENTS AND REPLACEMENTS, MOTOR (M16 - M19) - SORTER Install a new SCB.
  • Page 255: Error 30-936: Sorter Diverter #2 Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 30-936: SORTER DIVERTER #2 Failure 8F2919 Page 255 of 326 Messages LOCAL PANEL DICOM SEVERITY Sorter: Internal Hardware Sorter Diverter #2 failed to ELEC DOWN 1-Alert Failure operate (Alarm sounds) Description This error occurs when the SORTER DIVERTER ARM does not move, as determined by DIVERTER BIN 2 POSITION SENSOR (S21).
  • Page 256 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 256 of 326 CONNECTORS on the: Seat the CONNECTORS. • DIVERTER MOTOR • • DIVERTER MOTOR Install a new MOTOR. See the ADJUSTMENTS AND REPLACEMENTS, MOTOR (M16 - M19) - SORTER AY Install a new SCB.
  • Page 257: Error 30-937: Sorter Diverter #3 Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 30-937: SORTER DIVERTER #3 Failure 8F2919 Page 257 of 326 Messages LOCAL PANEL DICOM SEVERITY Sorter: Internal Hardware Sorter Diverter #3 failed to ELEC DOWN 1-Alert Failure operate (Alarm sounds) Description This error occurs when the SORTER DIVERTER ARM does not move, as determined by DIVERTER BIN 3 POSITION SENSOR (S22).
  • Page 258 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 258 of 326 CONNECTORS on the: Seat the CONNECTORS. • DIVERTER MOTOR • • DIVERTER MOTOR Install a new MOTOR. See the ADJUSTMENTS AND REPLACEMENTS, MOTOR (M16 - M19) - SORTER AY Install a new SCB.
  • Page 259: Error 30-938: Sorter Diverter #4 Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 30-938: SORTER DIVERTER #4 Failure 8F2919 Page 259 of 326 Messages LOCAL PANEL DICOM SEVERITY Sorter: Internal Hardware Sorter Diverter #4 failed to ELEC DOWN 1-Alert Failure operate (Alarm sounds) Description This error occurs when the SORTER DIVERTER ARM does not move, as determined by DIVERTER BIN 4 POSITION SENSOR (S23).
  • Page 260 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 260 of 326 CONNECTORS on the: Seat the CONNECTORS. • DIVERTER MOTOR • • DIVERTER MOTOR Install a new MOTOR. See the ADJUSTMENTS AND REPLACEMENTS, MOTOR (M16 - M19) - SORTER AY Install a new SCB.
  • Page 261: Error 32-154: Film Registration Nvram Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 32-154: FILM REGISTRATION NVRAM Failure 8F2919 Page 261 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Registration: Internal Film Registration: ELEC DOWN 0-Fatal Communications Failure microcontroller failed to read/ (Alarm sounds) write to NVRAM Description This error occurs when the MICROCONTROLLER detects an error when attempting to read or write from NVRAM.
  • Page 262 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 262 of 326 CONNECTORS on the PDB Seat the CONNECTORS. and FILM REGISTRATION AY Install a new FRB. See the ADJUSTMENTS AND REPLACEMENTS, FILM REGISTRATION BOARD (FRB) - FILM REGISTRATION Install a new PDB.
  • Page 263: Error 32-301: Film Registration Centering Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 32-301: FILM REGISTRATION Centering Failure 8F2919 Page 263 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Registration Failure Film Registration Centering FILM TRANS 1-Alert (Alarm sounds) Failure: Centering Home sensor S10 indicates Centering motor M7 failure Description This error occurs when the REGISTRATION CENTERING HOME SENSOR (S10) does not...
  • Page 264 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 264 of 326 For film jam Remove films. FILM CARTRIDGE Install a new FILM CARTRIDGE. BELTS Install new BELTS. ACCUMULATOR FILM Install a new SENSOR. See the ADJUSTMENTS AND DETECT SENSOR - use the REPLACEMENTS, SENSOR, ACCUMULATOR FILM...
  • Page 265: Error 32-302: Film Registration Clamping Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 32-302: FILM REGISTRATION Clamping Failure 8F2919 Page 265 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Registration Failure Film Registration Clamping FILM TRANS 1-Alert (Alarm sounds) Failure: Clamping Actuator Home sensor S11 indicates Clamping motor M8 failure Description This error occurs when the REGISTRATION CLAMP SENSOR (S11) does not detect...
  • Page 266 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 266 of 326 For film jam Remove films. CONNECTORS Seat the CONNECTORS. ACCUMULATOR FILM Install a new SENSOR. See the ADJUSTMENTS AND DETECT SENSOR - use the REPLACEMENTS, SENSOR, ACCUMULATOR FILM SERVICE TOOL...
  • Page 267: Error 32-304: Internal Temperature/Humidity Sensor Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 32-304: Internal Temperature/Humidity SENSOR Failure 8F2919 Page 267 of 326 Messages LOCAL PANEL DICOM SEVERITY None Internal Temperature/ NORMAL 3-Notify (No alarm) Humidity Sensor Failure Description This error occurs if the TEMPERATURE/RH SENSOR (S25) failed diagnostics. Result of Error None User Action...
  • Page 268: Error 32-931: Film Registration Communications Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 32-931: FILM REGISTRATION Communications Failure 8F2919 Page 268 of 326 Messages LOCAL PANEL DICOM SEVERITY Film Registration: Internal MCS failed to communicate ELEC DOWN 0-Fatal Communications Error with Film Registration after (Alarm sounds) multiple attempts Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but...
  • Page 269 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 269 of 326 CONNECTORS on the PDB Seat the CONNECTORS. and FILM REGISTRATION AY Install a new FRB. See the ADJUSTMENTS AND REPLACEMENTS, FILM REGISTRATION BOARD (FRB) - FILM REGISTRATION Install a new PDB.
  • Page 270: Error 32-932: Film Registration Communications Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 32-932: FILM REGISTRATION Communications Retry 8F2919 Page 270 of 326 Messages LOCAL PANEL DICOM SEVERITY None MCS failed to communicate NORMAL 3-Notify (No alarm) with Film Registration Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response.
  • Page 271: Error 33-154: Exposure Transport Nvram Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-154: EXPOSURE TRANSPORT NVRAM Failure 8F2919 Page 271 of 326 Messages LOCAL PANEL DICOM SEVERITY Exposure Transport: Internal Exposure Transport ELEC DOWN 0-Fatal Communications Failure microcontroller failed to read/ (Alarm sounds) write to NVRAM Description This error occurs when the MICROCONTROLLER detects an error when attempting to read or write from NVRAM.
  • Page 272 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 272 of 326 CONNECTORS on the PDB Seat the CONNECTORS. and ETB Install a new ETB. See the ADJUSTMENTS AND REPLACEMENTS, EXPOSURE TRANSPORT BOARD (ETB) - EXPOSURE TRANSPORT Install a new PDB.
  • Page 273: Error 33-311: Exposure Transport Entrance Feed Roller Failure272

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-311: EXPOSURE TRANSPORT ENTRANCE FEED ROLLER Failure 8F2919 Page 273 of 326 Messages LOCAL PANEL DICOM SEVERITY Exposure Transport: Entrance Exposure Transport Entrance FILM TRANS 1-Alert Failure Feed Roller Failure: Entrance (Alarm sounds) Roller Position sensor S14 indicates Exposure Transport Entrance Motor M10 failure...
  • Page 274 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 274 of 326 For film jam Remove films. CONNECTORS Seat the CONNECTORS. EXPOSURE ENTRANCE Install a new ETB. See the ADJUSTMENTS AND CLAMP SENSOR - use the REPLACEMENTS, EXPOSURE TRANSPORT BOARD SERVICE TOOL...
  • Page 275: Error 33-312: Exposure Transport Exit Feed Roller Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-312: EXPOSURE TRANSPORT EXIT FEED ROLLER Failure 8F2919 Page 275 of 326 Messages LOCAL PANEL DICOM SEVERITY Exposure Transport: Exit Exposure Transport Exit Feed FILM TRANS 1-Alert Failure Roller Failure: Exit Roller (Alarm sounds) Position sensor S15 indicates Exposure Transport Exit...
  • Page 276 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 276 of 326 For film jam Remove films. CONNECTORS Seat the CONNECTORS. EXPOSURE EXIT CLAMP Install a new ETB. See the ADJUSTMENTS AND SENSOR - use the SERVICE REPLACEMENTS, EXPOSURE TRANSPORT BOARD...
  • Page 277: Error 33-354: Exposure Transport Dsp Nvram Error

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-354: EXPOSURE TRANSPORT DSP NVRAM Error 8F2919 Page 277 of 326 Messages LOCAL PANEL DICOM SEVERITY Exposure Transport: Internal Exposure Transport DSP ELEC DOWN 0-Fatal Communications Failure failed to read/write to NVRAM (Alarm sounds) Description This error occurs when the DSP detects an error when attempting to read or write from...
  • Page 278 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 278 of 326 CONNECTORS on the PDB Seat the CONNECTORS. and ETB Install a new ETB. See the ADJUSTMENTS AND REPLACEMENTS, EXPOSURE TRANSPORT BOARD (ETB) - EXPOSURE TRANSPORT Install a new PDB.
  • Page 279: Error 33-355: Exposure Transport Dsp Communications Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-355: EXPOSURE TRANSPORT DSP Communications Failure 8F2919 Page 279 of 326 Messages LOCAL PANEL DICOM SEVERITY Exposure Transport: Internal MCS failed to communicate ELEC DOWN 0-Fatal Communications Failure with Exposure Transport DSP (Alarm sounds) after multiple attempts Description...
  • Page 280 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 280 of 326 CONNECTORS on the PDB Seat the CONNECTORS. and ETB Install a new ETB. See the ADJUSTMENTS AND REPLACEMENTS, EXPOSURE TRANSPORT BOARD (ETB) - EXPOSURE TRANSPORT Install a new PDB.
  • Page 281: Error 33-356: Exposure Transport Dsp Communications Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-356: EXPOSURE TRANSPORT DSP Communications Retry 8F2919 Page 281 of 326 Messages LOCAL PANEL DICOM SEVERITY None MCS failed to communicate NORMAL 3-Notify (No alarm) with Exposure Transport DSP Description This error occurs when the MCS tries to communicate with the DSP but receives no response.
  • Page 282: Error 33-380: Exposure Transport Motor Overcurrent Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-380: EXPOSURE TRANSPORT MOTOR Overcurrent Failure 8F2919 Page 282 of 326 Messages LOCAL PANEL DICOM SEVERITY Exposure Transport: Motor Exposure Transport Motor FILM TRANS 1-Alert Failure Overcurrent Failure: Drive (Alarm sounds) Motor overcurrent detected Description This error occurs when an overcurrent condition is detected on the DRIVE MOTOR (M12).
  • Page 283 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 283 of 326 CONNECTORS on the DRIVE Seat the CONNECTORS. MOTOR and ETB DRIVE MOTOR - use the Install a new MOTOR. See the ADJUSTMENTS AND SERVICE TOOL REPLACEMENTS, MOTOR, DRIVE (M12) -...
  • Page 284: Error 33-381: Exposure Transport Motor Power Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-381: EXPOSURE TRANSPORT MOTOR Power Failure 8F2919 Page 284 of 326 Messages LOCAL PANEL DICOM SEVERITY Exposure Transport: Motor Exposure Transport Motor FILM TRANS 1-Alert Failure Power Failure: 24V Power (Alarm sounds) Supply out of range Description This error occurs when the 24 V POWER SUPPLY is outside of operational limits.
  • Page 285 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 285 of 326 CONNECTORS on the ETB Seat the CONNECTORS. and PDB Install a new ETB. See the ADJUSTMENTS AND REPLACEMENTS, EXPOSURE TRANSPORT BOARD (ETB) - EXPOSURE TRANSPORT Install a new PDB.
  • Page 286: Error 33-385: Exposure Transport Motor Start Up Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-385: EXPOSURE TRANSPORT MOTOR Start Up Failure 8F2919 Page 286 of 326 Messages LOCAL PANEL DICOM SEVERITY Exposure Transport: Motor Exposure Transport Motor FILM TRANS 1-Alert Failure Start up Failure: Drive Motor (Alarm sounds) failed to reach proper speed after start up...
  • Page 287 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 287 of 326 CONNECTORS on the DRIVE Seat the CONNECTORS. MOTOR and ETB DRIVE MOTOR - use the Install a new MOTOR. See the ADJUSTMENTS AND SERVICE TOOL REPLACEMENTS, MOTOR, DRIVE (M12) -...
  • Page 288: Error 33-386: Exposure Transport Motor Speed Control Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-386: EXPOSURE TRANSPORT MOTOR Speed Control Failure 8F2919 Page 288 of 326 Messages LOCAL PANEL DICOM SEVERITY Exposure Transport: Motor Exposure Transport Motor FILM TRANS 1-Alert Control Failure Speed Control Failure: Drive (Alarm sounds) Motor repeatedly failed to maintain proper speed...
  • Page 289 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 289 of 326 CONNECTORS on the DRIVE Seat the CONNECTORS. MOTOR and ETB DRIVE MOTOR - use the Install a new MOTOR. See the ADJUSTMENTS AND SERVICE TOOL REPLACEMENTS, MOTOR, DRIVE (M12) -...
  • Page 290: Page Error 33-387: Exposure Transport Motor Speed Control Warning

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-387: EXPOSURE TRANSPORT MOTOR Speed Control Warning 8F2919 Page 290 of 326 Messages LOCAL PANEL DICOM SEVERITY None Exposure Transport Motor NORMAL 3-Notify (No alarm) Speed Control Warning: Drive Motor failed to maintain proper speed Description This error occurs when the DRIVE MOTOR (M12) does not maintain the proper speed when...
  • Page 291: Error 33-931: Exposure Transport Communications Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-931: EXPOSURE TRANSPORT Communications Failure 8F2919 Page 291 of 326 Messages LOCAL PANEL DICOM SEVERITY Exposure Transport: Internal MCS failed to communicate ELEC DOWN 0-Fatal Communications Failure with Exposure Transport after (Alarm sounds) multiple attempts Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but...
  • Page 292 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 292 of 326 CONNECTORS on the PDB Seat the CONNECTORS. and ETB Install a new ETB. See the ADJUSTMENTS AND REPLACEMENTS, EXPOSURE TRANSPORT BOARD (ETB) - EXPOSURE TRANSPORT Install a new PDB.
  • Page 293: Error 33-932: Exposure Transport Communications Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 33-932: EXPOSURE TRANSPORT Communications Retry 8F2919 Page 293 of 326 Messages LOCAL PANEL DICOM SEVERITY None MCS failed to communicate NORMAL 3-Notify (No alarm) with Exposure Transport Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response.
  • Page 294: Error 34-154: Pdb Nvram Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 34-154: PDB NVRAM Failure 8F2919 Page 294 of 326 Messages LOCAL PANEL DICOM SEVERITY PDB: Internal PDB microcontroller failed to ELEC DOWN 0-Fatal Communications Failure read/write to NVRAM (Alarm sounds) Description This error occurs when the MICROCONTROLLER detects an error when attempting to read or write from NVRAM.
  • Page 295: Error 34-207: Latch Unlock Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 34-207: LATCH Unlock Failure 8F2919 Page 295 of 326 Messages LOCAL PANEL DICOM SEVERITY Imager Unlock Mechanism Latch mechanism failed to COVER OPEN 1-Alert Failure open when requested (Alarm sounds) Description This error occurs when the user requests that the IMAGER unlock, but the LATCH fails to unlock as indicated by the LATCH POSITION SENSOR (S19).
  • Page 296 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 296 of 326 CONNECTORS on the: Seat the CONNECTORS. • LATCH MOTOR • LATCH SENSOR LATCH SENSOR - use the Install a new SENSOR. SERVICE TOOL LATCH MOTOR - use the Install a new MOTOR.
  • Page 297: Error 34-208: Latch Lock Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 34-208: LATCH Lock Failure 8F2919 Page 297 of 326 Messages LOCAL PANEL DICOM SEVERITY Imager Lock Mechanism Latch mechanism failed to COVER OPEN 1-Alert Failure close when requested (Alarm sounds) Description This error occurs when the user requests that the IMAGER lock, but the LATCH fails to lock as indicated by LATCH POSITION SENSOR (S19).
  • Page 298 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 298 of 326 CONNECTORS on the: Seat the CONNECTORS. • LATCH MOTOR • LATCH SENSOR LATCH SENSOR - use the Install a new SENSOR. SERVICE TOOL LATCH MOTOR - use the Install a new MOTOR.
  • Page 299: Error 34-931: Pdb Communications Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 34-931: PDB Communications Failure 8F2919 Page 299 of 326 Messages LOCAL PANEL DICOM SEVERITY PDB: Internal MCS failed to communicate ELEC DOWN 0-Fatal Communications Failure with PDB after multiple (Alarm sounds) attempts Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response to 5 attempts.
  • Page 300: Error 34-932: Pdb Communications Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 34-932: PDB Communications Retry 8F2919 Page 300 of 326 Messages LOCAL PANEL DICOM SEVERITY None MCS failed to communicate NORMAL 3-Notify (No alarm) with PDB Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response.
  • Page 301: Error 35-931: Processor Transport Communications Failure

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 35-931: PROCESSOR Transport Communications Failure 8F2919 Page 301 of 326 Messages LOCAL PANEL DICOM SEVERITY Processor Transport: Internal MCS failed to communicate ELEC DOWN 0-Fatal Communications Failure with Processor Transport (Alarm sounds) after multiple attempts Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but...
  • Page 302 DIAGNOSTICS Status Messages and Error Codes 30JUL07 Do the following: 8F2919 Page Check: If an existing problem: 302 of 326 CONNECTORS on the: Seat the CONNECTORS. • • Install a new PCB. See the ADJUSTMENTS AND REPLACEMENTS, PROCESSOR CONTROL BOARD (PCB) - THERMAL PROCESSOR Install a new PDB.
  • Page 303: Error 35-932: Processor Transport Communications Retry

    DIAGNOSTICS Status Messages and Error Codes 30JUL07 Error 35-932: Processor Transport Communications Retry 8F2919 Page 303 of 326 Messages LOCAL PANEL DICOM SEVERITY None MCS failed to communicate NORMAL 3-Notify (No alarm) with Processor Transport Description This error occurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response.
  • Page 304: Troubleshooting

    DIAGNOSTICS Troubleshooting 30JUL07 Section 3: Troubleshooting 8F2919 Page 304 of 326 System Problems Description Cause Action Cannot connect SERVICE SECURE LINK CLIENT Check the CERTIFICATE end LAPTOP with SERVICE SOFTWARE problem. date. TOOL Wrong type of network Use a CROSSOVER CABLE. CABLE.
  • Page 305: Power Supply Problems

    DIAGNOSTICS Troubleshooting 30JUL07 POWER SUPPLY Problems 8F2919 Page 305 of 326 Description or Possible Cause Action System will not power up. • Check source AC power at the wall. • Check the CIRCUIT BREAKERS. If either CIRCUIT BREAKER is open, reset. A cause might be a short in the POWER MODULE.
  • Page 306: Functions Of The Micro Board Leds, Test Points And Switches

    DIAGNOSTICS Troubleshooting 30JUL07 Functions of the MICRO BOARD LEDs, Test Points and 8F2919 Page SWITCHES 306 of 326 “Happy” LEDs on the MICRO BOARDS Important All the MICRO BOARDS on the MCS I C Bus include “HAPPY” LEDs which use blink ON/ OFF patterns to provide status information about the MICRO BOARDS.
  • Page 307: Power Distribution Board (Pdb)

    DIAGNOSTICS Troubleshooting 30JUL07 POWER DISTRIBUTION BOARD (PDB) 8F2919 Page 307 of 326 Table 1 LEDs and Test Points for the PDB Item Description D5 (LED) 24 V present. D6 (LED) 5 V present. D10 (LED) “Happy” LED. D13 (LED) DRAWER_HAZARD_24 present. D14 (LED) V24_HAZ present.
  • Page 308: Film Supply Board (Fsb)

    DIAGNOSTICS Troubleshooting 30JUL07 FILM SUPPLY BOARD (FSB) 8F2919 Page 308 of 326 Table 3 LEDs and Test Points for the FSB Item Description D5 (LED) “Happy” LED. DRAWER_HAZ_24 Test point for hazard 24 V. Test point for digital 5 V. MGND Test point for MOTOR ground.
  • Page 309: Film Registration Board (Frb)

    DIAGNOSTICS Troubleshooting 30JUL07 FILM REGISTRATION BOARD (FRB) 8F2919 Page 309 of 326 Table 5 LEDs and Test Points for the FRB Item Description D2 (LED) “Happy” LED. V24_HAZ Test point for hazard 24 V. Test point for digital 5 V. V3_3 Test point for digital 3.3 V.
  • Page 310: Exposure Transport Board (Etb)

    DIAGNOSTICS Troubleshooting 30JUL07 EXPOSURE TRANSPORT BOARD (ETB) 8F2919 Page 310 of 326 Table 7 LEDs and Test Points for the ETB Item Description D1 (LED) TX data - flashing indicates serial activity. D2 (LED) Enable SERVO MOTOR DRIVER - ON = DRIVER enabled. D3 (LED) RX data - flashing indicates serial activity.
  • Page 311 DIAGNOSTICS Troubleshooting 30JUL07 8F2919 Item Description Page TP10 Not used. 311 of 326 TP11 Test point for reset (active low reset pulse). TP12 Test point for CAP6_QEP12. TP13 Test point for ground TP14 Not used. TP15 Not used. TP16 Test point for CAP1QEP1 (filtered encoder phase A). TP17 Test point for CAP2QEP2 (filtered encoder phase B).
  • Page 312 DIAGNOSTICS Troubleshooting 30JUL07 8F2919 Item Description Page TP49 Test point for V24 hazard. 312 of 326 TP51 Test point for ground. TP52 Test point for 15 V DC. TP53 Test point for 12 V DC. TP55 Test point for MOTOR phase U HALL SENSOR. TP56 Test point for MOTOR phase V HALL SENSOR.
  • Page 313: Processor Control Board (Pcb)

    DIAGNOSTICS Troubleshooting 30JUL07 PROCESSOR CONTROL BOARD (PCB) 8F2919 Page 313 of 326 Table 8 LEDs and Test Points for the PCB Item Description DS1 (LED) MOTOR MICRO “happy” LED. DS2 (LED) K1 relay status (FLAZTBED high voltage return). ON = FLATBED high voltage return relay energized.
  • Page 314: Sorter Control Board (Scb)

    DIAGNOSTICS Troubleshooting 30JUL07 8F2919 Item Description Page Test point for 24 V hazard (V24_HAZ). 314 of 326 Test point for MOTOR ground (MGND). Reset SWITCH. SORTER CONTROL BOARD (SCB) Table 9 LEDs and Test Points for the SCB Item Description D1 (LED) “Happy”...
  • Page 315: Using The Logs

    DIAGNOSTICS Using the Logs 30JUL07 Section 4: Using the Logs 8F2919 Page 315 of 326 Types of Log The SERVICE TOOL includes 9 logs. • APPLICATION Log - This log records the “events” controlled and monitored by all system software applications and subsystems. •...
  • Page 316: Application Log

    DIAGNOSTICS Using the Logs 30JUL07 APPLICATION Log 8F2919 Page Using the APPLICATION Log 316 of 326 The Application Log and the other subsystem logs which can be “filtered” from the Application Log, are important troubleshooting resources. When an error occurs, you can search the logs to help locate the problem by checking the sequence of “events”...
  • Page 317: Log "Keying" Functions

    DIAGNOSTICS Using the Logs 30JUL07 • If you have the approximate time when the “event” you want to view occurred, select 8F2919 Page “View Log with time range” and enter a Start and End Time. 317 of 326 • If you want to view the log of an “event” when the “event” is occurring, select “View Active Log”.
  • Page 318 DIAGNOSTICS Using the Logs 30JUL07 Stop - Click [Stop] at any time during the time a log is loading to stop the loading 8F2919 Page process. 318 of 326 Options - Click [Options] to display a screen of the log “columns” you want to display. •...
  • Page 319: Using Advanced "Filtering

    DIAGNOSTICS Using the Logs 30JUL07 Using Advanced “Filtering” 8F2919 Page This section explains the use of the “filtering” function. For example: Modalities are not able to 319 of 326 select the IMAGER because the current number of “scp associations” has reached and remains at the limit (12).
  • Page 320: Contents Of The Log

    DIAGNOSTICS Using the Logs 30JUL07 Note in the retrieved data that only 5 “associations” are indicated “down”. 8F2919 Page A comparison of the up and down logs indicate that the “leonardo” modality continued to 320 of 326 add “associations” without closing down “associations” until the limit of 12 “associations” was reached.
  • Page 321: Log Levels

    DIAGNOSTICS Using the Logs 30JUL07 8F2919 MIM Subsystem Description Page 321 of 326 NETWORK This subsystem “negotiates” with the SCU for sending a job from the (DICOM) SCU across the network. It then makes a Film “Session”, sets up the Film “Box”...
  • Page 322 DIAGNOSTICS Using the Logs 30JUL07 From the SERVICE TOOL menu, select Configuration>System>Log Level. 8F2919 Page Click [Modify]. 322 of 326 From the drop-down menu for the subsystem, select the “Log Level” you want and click [Save]. Note The log levels, from lowest to highest are: •...
  • Page 323: Dicom Log

    DIAGNOSTICS Using the Logs 30JUL07 DICOM Log 8F2919 Page This is a subsystem log that includes all MIM Network “events” pulled from an Application 323 of 326 Log. Note that the Subsystem “column” includes only Network. See Sample Log 1 at the end of this section for descriptions of the “events”...
  • Page 324: Iis Log

    DIAGNOSTICS Using the Logs 30JUL07 Click [Save]. 8F2919 Page 324 of 326 Note Your LAPTOP must have LOCAL LOG VIEWER SOFTWARE to view the saved log. IIS Log The INTERNET INFORMATION SYSTEM (IIS) log records all “attempts” to connect to the Web site SERVER.
  • Page 325: Nt Event Log (System)

    DIAGNOSTICS Using the Logs 30JUL07 NT EVENT Log (System) 8F2919 Page This is a log of Windows “events” for system software. 325 of 326 PRINT DELIVERY Log This is a subsystem log that includes all DELIVERY and SDS SERVER SUBSYSTEM “events”...
  • Page 326: Publication History

    30JUL07 8F2919 CN0008637 8f2919.fm • New Publication • Rev A DryView is a trademark of Carestream Health. Kodak is a trademark of Kodak used under license. Printed in U.S.A. • 6800_Diagnostics_8f2919.fm Carestream Health, Inc. 150 Verona Street Rochester, NY 14608...