Tolerance Control For Differential Values - KERN AES Operation Manual

Analytical balances
Hide thumbs Also See for KERN AES:
Table of Contents

Advertisement

11.7.5 Tolerance control for differential values

• Entering 2 limits by weighing
1. How to enable the tolerance weighing function
[2.SEL.2] or [2.SEL.3], see chpt. 11.7.3.
2. For settings in menu see chpt. 11.7.3.
3. How to save settings / how to exit the menu
Press the S-key
Weighing balance is in tolerance mode, the trian-
gular tolerance tags are faded in.
4. How to set reference weight and limits:
Press the S-key for about 4 seconds until
[r. SET.] appears and then let go. The most re-
cently saved reference weight will be displayed
(flashing).
AES/AEJ-BA-e-0912
Setting for 2 limiting values:
Setting for differential value:
0.0000
z0000
M
g
g
63

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Kern aejAej 120-4mAej 220-4mAes 120-4Aes 220-4

Table of Contents