External Tare Command; Remote Control Instructions - KERN AES Operation Manual

Analytical balances
Hide thumbs Also See for KERN AES:
Table of Contents

Advertisement

14.5.2 External tare command

C1
C2
ASCII code
T
(SP)
54H
20H

14.5.3 Remote control instructions

C1
C2
Code
O
0
4FH
O
1
4FH
O
2
4FH
O
3
4FH
O
4
4FH
O
5
4FH
O
6
4FH
O
7
4FH
O
8
4FH
O
9
4FH
O
A
4FH
O
B
4FH
* Do not operate the PRINT-key whilst using these remote control commands
(would interrupt data transmission). If data transmission is interrupted disconnect
weighing balance briefly from mains.
AES/AEJ-BA-e-0912
Description
Value
Taring
None
Zeroing
Significance
30H
No data output
31H
Continuous data output
Continuous data output stable weighing
32H
values
Output for stable and instable weighing
33H
values after pressing PRINT key
Output for stable weighing value after
34H
previous relief of balance
One output for stable weighing value. No
35H
output for stable weighing values. Re-
newed output after stabilization
One output for stable weighing value.
36H
Continuous output for instable weighing
values.
Output of stable weighing values after
37H
pressing PRINT key
38H
Single immediate output*
39H
Single output after stabilization*
Single immediate output after a deter-
41H
mined interval*
Single immediate output after a deter-
42H
mined interval and a stable weighing
value*
Feedback
A00: Execution successful
E01: Error
E04: Taring / zero not possible, admissible
range exceeded.
A00:
E01:
E02:
Feedback
Free from
error
Error mes-
sage
Interval
error
(OA or OB
only)
81

Hide quick links: