9.2 Adjustment with external weight (only AEJ)

During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Condition: Menu setting [
Press the CAL key
Press the CAL-key again and the adjustment test
will take place automatically.
The difference between the saved value and the
measured value is displayed.
Adjustment protocols will be issued automatically
if you connect an optional printer, see chpt. 15
"printer mode". Whilst printing is in progress
[busy
] will be displayed.
Press any key; adjustment process will stop. The
balance returns to weighing mode.