KERN AES Operation Manual page 66

Analytical balances
Hide thumbs Also See for KERN AES:
Table of Contents

Advertisement

5. Press and hold the TARE-key until the display
used for entering the reference weight appears.
Each time the TARE-key is pressed, numbers will
run through from 0-9 and the decimal point.
0
Select the number to be changed using F key (the
active position flashes):
8. Confirm by pressing the S-key.
The reference weight will briefly be faded in, fol-
lowed by [L. SET].
6. Wait until the display starts flashing and the most
recently stored value is displayed.
Enter difference "reference weight – lower limit"
as described in step 5.
7. Confirm by pressing the S-key.
The value will briefly be faded in, followed by [H.
SET].
(This completes the entry provided only one limit
is set for tolerance control – menu setting
[23. P
.1] ).
1
8. Wait until the display starts flashing and the most
recently stored value is displayed.
Enter difference "reference weight – upper limit"
as described in step 5.
9. Save by pressing the S-key.
The balance returns to tolerance weighing mode.
Remove sample. From here evaluation takes place
whether the goods to be weighed are within the two
tolerance limits.
• Numeric entry of 3 or 4 limits
Menu setting [23.
P
For numeric entry of limits proceed as for entry of 2 limits. Instead of [L. SET] and
[H. SET] appears [L1. SET] - [L3. SET] or [L4. SET]. For this purpose repeat step 5
to 7 in each case.
66
1 ⇒
2 9
0
. 3] or [23.
P
. 4].
1
1
L. SET
H. SET
1
0.0000
AES/AEJ-BA-e-0912