5. Place the sample used to determine the reference
weight on the weighing pan.
6. Save by pressing the F-key. The reference weight
will briefly be faded in, followed by [L. SET].
7. Wait until the display starts flashing and the most
recently stored value is displayed.
9. Place sample for the lower (i.e. smaller) limiting
value on the weighing plate:
10. Save by pressing the F-key. The difference "ref-
erence weight – lower limit" will briefly be faded in,
followed by [H. SET].
11. Wait until the display starts flashing and the most
recently upper limit is displayed. Place sample for
the upper (i.e. larger) limiting value on the weigh-
ing plate:
12. Save by pressing the F-key. The difference "ref-
erence weight – upper limit" will briefly be faded
in, followed by the weight value of the sample.
The balance returns to tolerance weighing mode.
Remove sample. From here evaluation takes place
whether the goods to be weighed are within the two
tolerance limits.
Entering 3 or 4 limits by weighing
Menu setting [23.
P
For setting of limits proceed as for entry of 2 limits. Instead of [L. SET] and [H. SET]
appears [L1. SET] - [L3. SET] or [L4. SET]. For this purpose repeat step 5 to 7 in
each case.
Display of tolerance mark:
64
. 3] or [23.
P
. 4].
1
1
L. SET
0.0000
M
H. SET
1
0.0000
M
0.0000
AES/AEJ-BA-e-0912