9.7.1
Verification switch and seals
Prior to verification you will have move the calibration switch to verification position
(See direction of arrow, fig. 1). In this position the display shows a bracket around
the last display point.
After verification the balance is sealed at the indicated positions.
Seal of approval
Seal/rubber plug/verification switch
Position of
verification switch
forwards
backwards
10 Menu
In the menu the settings of the balance can be modified and functions can be acti-
vated. This way, the balance can be adjusted to individual weighing requirements.
The menu is sub-divided into "menu 1" and "menu 2".
AES/AEJ-BA-e-0912
Seal of approval
Fig. 1
Access to the verification switch is achieved
by removing the possibly existing seal (veri-
fication becomes invalid) and the rubber
plug.
Status
Balance is unlocked for the adjustment process, ad-
justment will be possible "ADJ" appears when the bal-
ance is switched on.
Verification position - Adjustment locked
Move verification switch to
verification position.