11.7.3 Enable function / settings in menu

Activate function
In weighing mode keep the F-key pressed down
until [FWnm] appears on the display. Release but-
ton. The first menu item showing the current set-
ting will be displayed.
Press the F-key and menu item [2 SEL] will
appear.
Select the desired setting by pressing the TARE-
key
[2] : Tolerance weighing
[ 3 ]: Tolerance weighing /Add-up
Select menu item "settings"
Press the F-key again
Press the TARE-key to select the desired setting,
see chpt. 10.1.4. From then on additional settings
can be selected each time the F-key is pressed.
How to change settings
Select the desired setting by pressing the TARE-
key, see chpt. 10.1.4.
Confirm by pressing the S-key.
How to save settings / how to exit the menu
Press the S-key
Weighing balance is in tolerance mode, the trian-
gular tolerance tags are faded in.
AES/AEJ-BA-e-0912
Fonm
The first menu item for setting
the tolerance marker ap-
pears.
0.0000
g
57