• Numeric entering of 2 limits
Example:
Weight value
Input
1. How to enable the tolerance weighing function
[2.SEL.2] or [2.SEL.3], see chpt. 11.7.3.
2. For settings in menu see chpt. 11.7.3.
3. How to save settings / how to exit the menu
Press the S-key
Weighing balance is in tolerance mode, the trian-
gular tolerance tags are faded in.
4. How to set reference weight and limits:
Press the S-key for about 4 seconds until
[r. SET.] appears and then let go. The most re-
cently saved reference weight will be displayed
(flashing).
AES/AEJ-BA-e-0912
Reference weight
100.0000 g
100.0000 g
Lower limit
97.0000 g
-3.0000 g
Setting for 2 limiting values:
Setting for differential value:
M
Upper limit
105.0000 g
5.0000 g
0.0000
g
z0000
g