Teledyne 401 Instruction Manual
Teledyne 401 Instruction Manual

Teledyne 401 Instruction Manual

O3 calibrator
Table of Contents

Advertisement

Quick Links

INSTRUCTION MANUAL
MODEL 401
O
CALIBRATOR
3
© TELEDYNE INSTRUMENTS
ADVANCED POLLUTION INSTRUMENTATION DIVISION
(T-API)
6565 NANCY RIDGE DRIVE
SAN DIEGO, CA 92121-2251
TOLL-FREE: 800-324-5190
FAX: 858-657-9816
TEL: 858-657-9800
E-MAIL: api-sales@teledyne.com
WEB SITE: www.teledyne-api.com
01124
REV. J1
Copyright 2004 API Inc.
12/20/04

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne 401

  • Page 1 INSTRUCTION MANUAL MODEL 401 CALIBRATOR © TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API) 6565 NANCY RIDGE DRIVE SAN DIEGO, CA 92121-2251 TOLL-FREE: 800-324-5190 FAX: 858-657-9816 TEL: 858-657-9800 E-MAIL: api-sales@teledyne.com WEB SITE: www.teledyne-api.com 01124 REV. J1 Copyright 2004 API Inc. 12/20/04...
  • Page 2 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol.
  • Page 3: Table Of Contents

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 TABLE OF CONTENTS SAFETY MESSAGES ..................II TABLE OF CONTENTS ..................III LIST OF FIGURES ..................... VII LIST OF TABLES....................VIII 1 INTRODUCTION .................... 1-1 1.1 P ............................1-1 REFACE 1.2 W...
  • Page 4 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 2.1.2 O Analog Output........................2-1 2.1.3 E ROM Backup Of Software Configuration ................2-1 2.1.4 Adaptive Filter ........................2-1 2.1.5 Data Acquisition ........................2-2 2.1.6 RS-232 Interface ........................2-2 2.1.7 Password Protection.......................
  • Page 5 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 5.2.1 D/A Output ..........................5-2 6 HANDLING WARNINGS ................6-1 7 RS-232 COMMUNICATIONS................. 7-1 7.1 DAS R ..........................7-2 EPORTING 7.2 W ............................7-3 ARNINGS 7.3 S ........................... 7-4...
  • Page 6 .................. 11-3 PARE ARTS NITS Appendix A LIST OF AVAILABLE MODEL 401 OPTIONS ..........A-1 Appendix B RECOMMENDED ZERO AIR SYSTEM............. B-1 Appendix C REFERENCES ....................... C-1 Appendix D SOFTWARE MENU TREES ................D-1 Appendix E ELECTRICAL SCHEMATIC INDEX ..............E-1...
  • Page 7: List Of Figures

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 LIST OF FIGURES 1-1: M 401 O ............1-11 IGURE ODEL ZONE ALIBRATOR LOCK IAGRAM 1-2: M401 R ..............1-12 IGURE ANEL LECTRICAL ONNECTIONS 1-3: R (CE V ) .......
  • Page 8 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 LIST OF TABLES 1-1: T ........................1-21 ABLE ALUES 2-1: P ....................... 2-3 ABLE ASSWORD EVELS 2-2: S ........................2-5 ABLE YSTEM ODES 2-3: T ......................2-6 ABLE...
  • Page 9: Introduction

    Teledyne API a very user-friendly system. We hope you will not experience any problems with the Teledyne API Model 401 but if you do, the built-in tests and diagnostics should allow you to quickly and easily find the problem. In addition, our full time customer service department is always available to answer your questions.
  • Page 10: Principle Of Operation

    The warranties set forth in this section shall be of no force or effect with respect to any Product: (i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used in any manner other than in accordance with the instruction provided by Teledyne API or (iii) not properly maintained.
  • Page 11 Photometric Calibrator Instruction Manual, 01124, Rev. J1 NOTE Model 401 Photometric Calibrator is designed to be a dedicated calibrator. The M401 Photometer should not be used as a monitor. The Beer-Lambert equation, shown below, calculates the concentration of ozone from the ratio of light intensities.
  • Page 12: Interferent Rejection

    It should be noted that the UV absorption method for detecting ozone is subject to interference from a number of sources. The Model 401 has been successfully tested for its ability to reject interference from sulfur dioxide, nitrogen dioxide, nitric oxide, water, and meta-xylene.
  • Page 13: Specifications

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 1.4 Specifications Photometer Ranges User selectable to any full scale range from 100 ppb to 20,000 ppb Zero Noise < 0.3 ppb (rms) Span Noise < 0.5% of reading (rms) Lower Detectable Limit <...
  • Page 14: Noise (Ppb Rms)

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 1.4.1 Noise (PPB RMS) The standard deviation of reported concentration calculated over 25 samples times 7 runs minimum, N-1 weighted. 1.4.2 Lower Detectable Limit (LDL) PPB LDL is defined as twice the (RMS) noise at the lower end of the most sensitive operating range.
  • Page 15: Flow Rates

    3. When installing the Model 401, allow a minimum of 4 inches (100mm) of clearance at the back of the instrument and 1 inch (25mm) of clearance on each side for proper ventilation.
  • Page 16 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 NOTE See Figure 1-4 for rear panel pneumatic connections. Span gas should only come into contact with PTFE or glass. Leak check all fittings with soap solution. Maximum pressure for leak check is 15 PSIG.
  • Page 17 In order to increase pressure, adjust knob clockwise as needed while monitoring the O flow test measurement display. In general, the O generator flow should be at least 1 L/MIN in addition of total flow demands for the Model 401 (800 cc/min) and remote analyzers. NOTE Adjustment of the O generator flow is to be performed by qualified maintenance personnel only.
  • Page 18 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 14. Generate the Zero Air. a. Press ZERO from the main menu and enter the password (512). b. Press STBY to terminate or press O GN to continue to generate O span gas.
  • Page 19: Figure 1-1: Model 401 O

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure 1-1: Model 401 Ozone Calibrator Block Diagram 1-11...
  • Page 20: Figure 1-2: M401 Rearp

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure 1-2: M401 Rear Panel Electrical Connections 1-12...
  • Page 21: Figure 1-3: Rear Panele

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure 1-3: Rear Panel Electrical and Pneumatic Connections (CE Version) 1-13...
  • Page 22: Electrical And Pneumatic Connections

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 1.6 Electrical And Pneumatic Connections 1.6.1 Electrical Connections Output #1 Ozone concentration - chart recorder Output #2 Ozone concentration - DAS Output #3 Test function analog output Input #4 Zero valve request...
  • Page 23: Figure 1-4: Rs-232 Pina

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure 1-4: RS-232 Pin Assignments 1-15...
  • Page 24: Pneumatic System

    Photometric Calibrator Instruction Manual, 01124, Rev. J1 1.6.3 Pneumatic System The Model 401 Photometric Calibrator is equipped with a compression pump to deliver up to 5 L/MIN for ozone generator system. It is also equipped with a second pump capable of pulling 800 cc/min across a critical flow orifice.
  • Page 25: Figure 1-5: M401 Calibrator

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure 1-5: M401 Calibrator Pneumatic Connections 1-17...
  • Page 26: Figure 1-6: M401 Rearp

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure 1-6: M401 Rear Panel (External Dry Zero Air) 1-18...
  • Page 27: Figure 1-7: M401 Calibrator

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure 1-7: M401 Calibrator Chassis Layout 1-19...
  • Page 28: Operation Verification

    12. If the TEST functions are within the limits given above the instrument should function correctly. If there is a problem please read the manual and check your set-up. The Model 401 is now ready for calibration, see Section 3.
  • Page 29: Final Test And Calibration Values

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 1.7.1 Final Test And Calibration Values Table 1-1: Test Values Test Parameter Displayed Value Units TIME HH:MM:SS DCPS BOX TEMP DEG C OFFSET SLOPE REG PRESSURE IN-HG-A GEN TMP...
  • Page 30: Options

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 1.8 Options 1.8.1 Rack Mount With Slides This option, including slides and rack mounting ears, permits the Analyzer to be mounted in a standard 19" wide x 30" deep RETMA rack.
  • Page 31: Operation

    2.1.3 E ROM Backup Of Software Configuration The Teledyne API Photometric Calibrator has few DIP switches or jumpers that need to be set by the operator. Configuration of the calibrator is done under software control and the configuration options are stored in electrically erasable (E ) ROM.
  • Page 32: Data Acquisition

    Photometric Calibrator Instruction Manual, 01124, Rev. J1 2.1.5 Data Acquisition The Teledyne API Photometric Calibrator contains a built-in data acquisition system which keeps track of the average O readings and the last 100 averages. This data is made available to other systems via the RS-232 interface.
  • Page 33: Table 2-1: Passwordl

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 2-1: Password Levels Password Level Functions Allowed No password required TEST, MSG, CLR, SETUP-DAS-VIEW Operator (512) ZERO, 03GN, ASEQ Setup (818) SETUP-EVNT, SETUP-BCAL, SETUP-COMM, SETUP-MISC-O3 GEN, SETUP-MISC-DARK,...
  • Page 34: Front Panel

    MENU field which occupies the entire bottom line of the display. A typical display is shown in Figure 2-1. Figure 2-1: Model 401 Front Panel The mode field indicates the current mode of the Analyzer. Usually, it shows "STBY ZS ", indicating that the instrument is in the standby mode and that automatic zero/span checking is enabled.
  • Page 35: Table 2-2: System Modes

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 2-2: System Modes Mode Meaning STBY xxxx (1) Standby mode (Photometer is operating normally) ZERO CAL x (2) Zero check in progress GN CAL x (2) Span check in progress...
  • Page 36: Table 2-3: Testm

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 2-3: Test Measurements Test Message Meaning TIME=xx:xx:xx Current time-of-day (HH:MM:SS) SET =xxxx PPB Ozone generator set concentration MEAS=xxxx MV Current measure channel in mV REF =xxxx MV...
  • Page 37: Programmable Pushbuttons

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 2-4: Warning Messages Warning Message Meaning SYSTEM RESET Issued whenever Analyzer is powered on RAM INITIALIZED RAM was erased (including DAS reports) ANA LAMP WARNING Analyzer lamp < 40000 OR >= 80000 counts...
  • Page 38: Figure 2-2: Illustration

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure 2-2: Illustration Of Normal Display If TEST is pushed, the upper center display cycles through the menu of test parameters, e.g. Sample flow (see Table 2-3). If STBY is pushed, the calibrator is in stand-by mode.
  • Page 39: Status Leds

    Photometric Calibrator Instruction Manual, 01124, Rev. J1 2.2.3 Status LEDs The three status LEDs to the right of the display indicate the general status of the Model 401 Photometric Calibrator. The green SAMPLE LED indicates the sampling status. The yellow CAL LED indicates the calibration status.
  • Page 40 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 INTENTIONALLY BLANK 2-10...
  • Page 41: Calibration

    The M401 is designed to be used as a dedicated ozone calibrator with built-in photometer. Zero/Span gas generation and calibration of Model 401 are discussed in Section 3. The emphasis in Section 3 is on the operation of the buttons and the internal adjustments they make in the instrument.
  • Page 42: Automatic Zero/Span (Or Multipoint) Check

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 3.3 Automatic Zero/Span (or Multipoint) Check Automatic zero/span gas supply must be enabled in the setup mode. There are five parameters that affect zero/span checking: the zero and span duration, the time of day for calibration, the time shift, the zero and multipoint calibration switch, and the feedback mode.
  • Page 43: Summary Of Front Panel Calibration Controls

    Photometer Zero Calibration 1. Turn the power on for both Model 401 and Model 400 (as master standard) and allow them to stabilize for minimum of one hour. 2. Start to generate the zero gas by pressing "zero" button from the main menu and enter the password (512).
  • Page 44: Remote Zero/Span Check (Contact Closure)

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 3. Allow to sample zero gas for approximately 10 minutes. 4. Follow sequence of SETUP, BCAL, and ZERO. Then select "yes" to adjust zero. NOTE: in case of using external dry zero air, check if dry zero air is supplied into the rear panel port (see Figure 1-5 for pneumatic connections) at about 30 - 40 PSI pressure.
  • Page 45: Remote Zero/Span Check Or Adjustment (Rs-232)

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 NOTE Teledyne API recommends that contact closures remain closed at least 10 minutes to allow for an accurate average zero or span value to be established. 3.8 Remote Zero/Span Check Or Adjustment (RS-232) Besides Z/S checking from the front panel, automatic Z/S checking, and remote Z/S checking via the contact closure inputs, the Analyzer can also be checked via the RS-232 interface.
  • Page 46 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 INTENTIONALLY BLANK...
  • Page 47: Setup Mode

    WAIT. Enter the duration between 1 - 20 minutes for the zero check and then press ENTR. similarly enter the duration for span check and press ENTR. NOTE Teledyne API recommends that the durations are set at least 10 minutes to allow for an accurate and reliable ZERO or SPAN value to be established.
  • Page 48: Examining The Ozone Formula Slope And Offset

    The current value of the ozone reading that is displayed on the front panel and output on the D/A terminals on the back panel is computed as follows: 1. The Model 401 Analyzer switches into reference mode. 2. The Analyzer waits 2 seconds to purge the sample tube.
  • Page 49: Setting The O Generation Mode

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 4.5 Setting the O Generation Mode There are three modes of feedback operation that can be selected for the ozone generator. The mode can be changed by pressing SETUP-MISC-O3-GEN-MODE. The three modes are shown below.
  • Page 50: Adjusting The Clock Speed

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 4.8 Adjusting The Clock Speed In order to compensate for clocks which run a little bit fast or slow, there is a variable called SETUP-MISC-CLK-ADJ. This variable is set to the number of seconds per day by which to speed up or slow down the clock.
  • Page 51: Setting The Analyzer I.d

    4.14 Setting The Fault LED Timeout The Model 401 can be set up so that the red fault LED on the front panel turns off automatically if no warning conditions persist. Any warning messages will still be displayed, however. To set the fault LED timeout, press SETUP-MISC-MORE-FILT and enter a value of from 0 to 300 seconds.
  • Page 52: Software Configuration

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 4.15 Software Configuration The software configuration can be displayed by entering the button sequence SETUP-CFG- LIST. For example the M401 could display: "O3 CALIBRATOR" "SBC40 CPU" Stating that the instrument was an Ozone Calibrator using the SBC40 computer. This feature is useful for showing any special features that are present in the currently installed PROM.
  • Page 53: Table 4-1: Setup Variables

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 4-1: Setup Variables Button Sequence Function Default Limits Zero cal. duration 15 min. 1-20 min. SETUP-EVNT-WAIT Span cal. duration 15 min. 1-20 min. SETUP-EVNT-TIME Cal. time-of-day 23:30...
  • Page 54 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 INTENTIONALLY BLANK...
  • Page 55: Diagnostics

    Viewing these test measurements does not interfere with the operation of the Model 401 or the ozone reading of the photometer in any way, so they may be viewed freely. Additionally, the values of most TEST functions can output as an analog voltage at the instrument's rear panel (see Figure 1-2).
  • Page 56: D/A Output

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 5-1: Test Channel Output Test # Name Purpose NONE No output O3 DETECT Outputs O detector REF DETECT Outputs IZS ref. det. SAMP PRESS Outputs sample press.
  • Page 57: Handling Warnings

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 6 HANDLING WARNINGS When a system warning occurs, a warning message is displayed and the FAULT LED blinks. A warning indicates that something in the system needs to be checked or adjusted. Failure by the operator to respond to a warning may result in poor system performance and/or less accurate data acquisition.
  • Page 58 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 INTENTIONALLY BLANK...
  • Page 59: Communications

    The uniform nature of the output messages makes it easy for a host computer to parse them. Input messages to the Model 401 have a format which is similar to that for output messages: "X COMMAND<CRLF>"...
  • Page 60: Das Reporting

    "R DDD:HH:MM IIII RANGE=xxxx O3=xxxx PPB SAMPLES=xx<CRLF>" Whenever the Model 401 is in calibration or diagnostics mode, no O readings are included in the average. Thus, it is possible for an average to contain 0 samples. If the number of samples in an average is 0, then "XXXX"...
  • Page 61: Warnings

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 7.2 Warnings Whenever a warning message is displayed on the display, it is also sent to the RS-232 output. See Table 2-4 for a list of the warning messages. These messages are very helpful when trying to track down a system problem and for determining whether or not DAS average data is actually valid.
  • Page 62: Status/Control

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 7-2: Warning Message Clear Commands Command Warning Message Cleared "W WSYSRES<CRLF>" SYSTEM RESET "W WRAMINIT<CRLF>" RAM INITIALIZED "W WALMPINT<CRLF>" ANA LAMP WARNING "W WO3LMPHLT<CRLF>" O3 GEN LAMP SHUTDOWN "W WO3REFDET<CRLF>"...
  • Page 63: Table 7-4: Controlc

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 To do a remote adjustment via the RS-232 interface, the host computer should issue a message with the following format: "C COMMAND<CRLF>" The commands are summarized in Table 7-4.
  • Page 64: Diagnostics

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 7.4 Diagnostics The diagnostics mode can be entered from the RS-232 port as well as from the front panel. The diagnostics commands available are listed below. NOTE The diagnostics mode may only be entered via the RS-232 port when the calibrator is in stand by mode.
  • Page 65: Test Measurements

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 7-6: Special Diagnostic Commands Command Function "D SYS-RESET<CRLF>" Reset Calibrator (same as power off/on) "D RAM-RESET<CRLF>" Reset Calibrator and erase RAM "D EE-RESET <CRLF>" Reset Calibrator and erase RAM & EEPROM Whenever the diagnostic mode is entered or exited, a report is issued to the RS-232 output.
  • Page 66: Viewing And Modifying Variables

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 7-8: Test Measurement Request Commands Command Test Measurement "?<CRLF>" RS-232 HELP screen "T LIST<CRLF>" Summary of all Tests "T O3MEAS<CRLF>" Current O meas counts "T O3REF<CRLF>" Current O reference counts "T SPRESS<CRLF>"...
  • Page 67 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 For example, to see the analyzer lamp temperature set point, the host computer would issue the command: "V ALAMP_SET<CRLF>" and the CPU would respond with something like: "V...
  • Page 68: Table 7-9: Rs-232 Variable

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 7-9: RS-232 Variable Names Var. Name Button Sequence Legal Values AUTO-TIME SETUP-IZSC-TIME 00:00-23:59 AUTO-SHIFT SETUP-IZSC-SHFT -60 - +60 TIME-ZERO SETUP-IZSC-WAIT 1-20 TIME-SPAN SETUP-IZSC-WAIT 1-20 O3-SPAN SETUP-IZSC-SPAN 0-9000...
  • Page 69: Adjustments

    The operations outlined in this chapter are to be performed by qualified maintenance personnel only. All adjustments to the Model 401 are easy to make. Pots and test points are readily accessible without removing any components. Figure 1-7 is a plan view of the Model 401 Photometric Calibrator showing all the major components.
  • Page 70: Figure 8-1: Electricalb

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure 8-1: Electrical Block Diagram...
  • Page 71: A/D - D/A Calibration Procedure

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 8.2 A/D - D/A Calibration Procedure Due to the stability of modern electronics, this procedure should not have to be performed more than once a year or whenever a major sub-assembly is exchanged or whenever analog output voltage range is changed .
  • Page 72: Output Voltage Range Changes

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Next, to automatically calibrate all the DACs, press SETUP-MISC-D/A-CAL-DAC. The display will show the percent completion as the analyzer goes through the procedure. You must calibrate the ADC before calibrating the DACs because the ADC is used during the DAC calibration procedure.
  • Page 73: Figure 8-2: Flow Andp

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure 8-2: Flow And Pressure Readout Adjustment...
  • Page 74: Dc Power Supply

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 8.5 DC Power Supply Overall performance of the DC power supply may be checked by observing the value displayed during test DCPS. If this value, a composite of the five (5) regulator outputs, deviates by more than 10% from the value recorded in Table 1-1 of this manual under Test Values, the outputs of the individual regulators should be measured.
  • Page 75: Status Lines

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 8.7 Status Lines Table 8-1: The Calibrator Has The Following Status Conditions Pin Pair Output # Status Condition High/Low 1, 2 ZERO CAL In Zero Calibrate 3, 4...
  • Page 76: Ozone Generator Lamp Setup

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 CAUTION UV light present. Do not pull the lamp from the optical bench assembly. 3. Adjust the UV Detector Pre-Amp gain as follows: a. Remove the access cap on the Detector cover at the front end of the optical bench, and REF reading on the display is 4500 mV ±...
  • Page 77: Ozone Generator Calibration

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 8.10 Ozone Generator Calibration The ozone generator can be calibrated against the analyzer calibration by using the analytical section of the M401 to determine the ozone generator's output. Calibration of the generator allows the operator to enter the desired calibration concentration directly in PPB's.
  • Page 78: Bench Feedback Option

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 To view the current dark offset, press SETUP-MISC-DARK-VAL. Press EXIT when finished. No password is required to view the dark offset, only to change it. 8.12 Bench Feedback Option The Bench Feedback Option uses the Photometer to control the lamp drive current.
  • Page 79: Troubleshooting

    9.1 Overview The Model 401 Photometric Calibrator has been designed to rapidly detect possible problems and allow their quick evaluation and repair. During operation, the calibrator continuously performs self-check diagnostics and provides the ability to monitor the key operating parameters of the instrument without disturbing monitoring operations.
  • Page 80: Checking The Power Sub-Systems

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 If these actions all occur, it is probable that the Calibrator's Power Supplies, CPU, and Display are working properly. If any of these actions fail to occur, power and/or CPU operation should be checked as follows.
  • Page 81: Checking The Cpu And Display

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 9.2.2 Checking The CPU And Display When the analyzer is turned on, the front panel display should energize and the green "Sample" LED should light. If proper DC power is present (see Section 9.2.1), the absence of these actions will usually indicate either a CPU or Display failure.
  • Page 82: Table 9-1: Warningm

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 9-1: Warning Messages Warning Message Meaning Corrective Action ANA LAMP The O REF value is greater than 5000 mV Check and adjust Source Lamp WARNING or less than 2500 mV.
  • Page 83: Troubleshooting Using Test Function Values

    Photometric Calibrator Instruction Manual, 01124, Rev. J1 9.4 Troubleshooting Using Test Function Values The Model 401 Photometric Calibrator provides the capability to display, on operator demand, the values of Test Functions which allow the observation of key analyzer operating parameters.
  • Page 84: Troubleshooting Dynamic Problems

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 9-2: Test Function Values (Continued) Test Corrective Action for Meaning Acceptable Values Function Unacceptable Values SAMPLE Sample mass flow rate. 700-900 cc./min Check for pneumatic system FLOW problems.
  • Page 85: Noisy Or Unstable Readings At Zero

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 9.5.1 Noisy Or Unstable Readings At Zero 1. Check for leaks in the pneumatic system as described in Section 10.3. 2. Confirm that the Zero gas is free of Ozone.
  • Page 86: Analog Outputs Do Not Agree With Front Panel Readings

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 9.5.4 Analog Outputs Do Not Agree With Front Panel Readings 1. Confirm that the DAC offset (SETUP-MISC-D/A-OFFS) is set to zero. 2. Perform a DAC/ADC adjustment and Dark Signal adjustment by following the procedure described in Sections 8.2 and 8.3.
  • Page 87: Troubleshooting Individual Sub-Assemblies And Components

    2. Flow is greater than zero, but is too low, and/or unstable. 3. Flow is too high. Figure 1-1 in this Manual provides a schematic diagram of the Flow in a Model 401. Flow Is Zero 1. Confirm that the sample pump (sample flow) and/or O...
  • Page 88: Troubleshooting Temperature Problems

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 9.6.2 Troubleshooting Temperature Problems The Model 401 has been designed to operate at ambient temperatures between 5 C and 40 C. As a first step in troubleshooting temperature problems, confirm the ambient temperature is within this range and that the air inlets slots on the sides of the cover and the fan exhaust on the rear panel are not obstructed.
  • Page 89: Checking The V/F Card

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Chassis Temperature Turn the analyzer off and remove the DC Power Supply Board. Measure across Motherboard J21 pins A30 and C30. If thermistor resistance(s) are within the proper range, check the temperature linearization circuits on the DC Power Supply Board as described in Section 9.6.4.
  • Page 90: Table 9-3: V/F Boardj

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 2. Confirm that all jumpers on the V/F board are set properly, as follows: Table 9-3: V/F Board Jumpers - Factory Settings Factory Set Jumpers Jumper Setting 3-4 (0-5V)
  • Page 91: Checking The Dc Power Supply Board

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 9.6.4 Checking The DC Power Supply Board A schematic and physical diagram of the DC Power Supply Board are shown on Drawings 00015 and 00016 in Appendix D.
  • Page 92: Checking The Pneumatic Sensor Board

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Four Temperature linearization circuits are contained on the DC Power Supply board. The outputs of these circuits can be checked by measuring the voltages at test points on the board as follows: 30 °...
  • Page 93: Table 9-5: Uv Sourcel

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 9-5: UV Source Lamp And Detector Diagnostics UV Source Lamp and Detector Diagnostics REF Value Meaning The Source Lamp and Detector are operating, but 4700 mV to 5000 mV adjustment is required.
  • Page 94: Checking Main Switching Valve

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 NOTE In cold, ambient conditions, it may require 5 to 15 minutes of warm- up before the source lamp initially fires. Be sure to wait for this period before troubleshooting the lamp/detector.
  • Page 95 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 WARNING Hazardous voltage present - use caution. 1. Unplug the cable connector at J1 on the Lamp Power Supply and confirm that +24VDC is present between Pins 1 and 2 on the cable connector. If this voltage is incorrect, check the Power Supply Module as described in Section 9.6.4.
  • Page 96: Warranty/Repair Questionnaire

    IF POSSIBLE, PLEASE INCLUDE A PORTION OF A STRIP CHART PERTAINING TO THE PROBLEM. CIRCLE PERTINENT DATA. 4. THANK YOU FOR PROVIDING THIS INFORMATION. YOUR ASSISTANCE ENABLES TELEDYNE API TO RESPOND FASTER TO THE PROBLEM THAT YOU ARE ENCOUNTERING TELEDYNE API CUSTOMER SERVICE EMAIL: api-customerservice@teledyne.com...
  • Page 97: Routine Maintenance

    10.1 Maintenance Schedule Table 10-1 shows a typical maintenance schedule for the Model 401. Please note that in certain environments (i.e. dusty, very high ambient pollutant levels) some maintenance procedures may need to be performed more often than shown.
  • Page 98: Leak Checking

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 10.2 Leak Checking There are two methods of leak checking: 1. By vacuum: This is the simplest method but it does not show the location of a leak.
  • Page 99: Changing The Prom

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 10.4 Changing The Prom 1. Locate the CPU card by referring to Figure 1-7. 2. Remove the screws that hold the CPU card (SBC40 printed on the lever) top corner to the V/F card, then remove the card from the STD-BUS backplane.
  • Page 100 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 INTENTIONALLY BLANK 10-4...
  • Page 101: Model 401 Spare Parts And Expendables Kits

    EXPENDABLES KITS 11.1 Spare Parts For CE Mark Units NOTE Use of replacement parts other than those supplied by Teledyne API may result in non-compliance with European Standard EN 61010-1. Table 11-1: Spare Parts For CE Mark Units Part No.
  • Page 102 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 11-1: Spare Parts For CE Mark Units (Continued) Part No. Description 01112 M401 Expendables Kit 01509 UV Lamp Power Supply 01916-01 Rear Panel Connector Board 01930 Keyboard...
  • Page 103: Spare Parts For Non-Ce Mark Units

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 11.2 Spare Parts For Non-CE Mark Units Table 11-2: Spare Parts For Non-CE Mark Units Part No. Description 00015 Power Supply Board 00276-0401 CPU Board 00402-03 Sensor Board...
  • Page 104 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Table 11-2: Spare Parts For Non-CE Mark Units (Continued) Part No. Description FM004 Flow Meter, 0-1000 CC FM05 Flow Restrictor (O Generator) FM06 Flow Restrictor (Zero Reference) HW020...
  • Page 105: Table A-1: Availablem

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Appendix A LIST OF AVAILABLE MODEL 401 OPTIONS Table A-1: Available Model 401 Options Option RS-232 Status Outputs Rack Mount and Slides Rack Mount Only...
  • Page 106 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 INTENTIONALLY BLANK...
  • Page 107: Appendix B Recommended Zero Air System

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Appendix B RECOMMENDED ZERO AIR SYSTEM The following is from the TECHNICAL ASSISTANCE DOCUMENT FOR THE CALIBRATION OF AMBIENT OZONE MONITORS - Richard J. Paur and Frank F. McElroy (1979).
  • Page 108 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 INTENTIONALLY BLANK...
  • Page 109: Appendix C References

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Appendix C REFERENCES 1. Calibration of Ozone Reference Methods, Code of Federal Regulations, Title 40, Part 50, Appendix D. 2. Technical Assistance Document for the Calibration of Ambient Ozone Monitors, EPA publication available from EPA, Department E (MD-77), Research Triangle Park, N.C.
  • Page 110 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 INTENTIONALLY BLANK...
  • Page 111: Appendix D Software Menu Trees

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Appendix D SOFTWARE MENU TREES Figure A-1: Model 401 Sample Software Menu Tree...
  • Page 112: Figure A-2: Model 401 S

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Figure A-2: Model 401 Setup Software Menu Tree...
  • Page 113: Appendix E Electrical Schematic Index

    Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 Appendix E ELECTRICAL SCHEMATIC INDEX Table E-1: Electrical Schematic Index Part No. Name 00015 Power Supply Board Assembly 00016 Power Supply Board Diagram 00402 Pneumatic Sensor Assembly 00403...
  • Page 114 Teledyne API Model 401 O Photometric Calibrator Instruction Manual, 01124, Rev. J1 INTENTIONALLY BLANK...