History:power:cmaximum? <Supply - HP E8402A User And Service Manual

Vxi c-size mainframe
Table of Contents

Advertisement

HISTory:POWer:CMAXimum? < supply >
Parameters
Returned Data
Comments
Error Conditions
Reset Condition
Related
Commands
68 Programming the Enhanced Monitor
HISTory:POWer:CMAXimum?
the maximum wattage measured by the Enhanced Monitor for the <supply> since
power-on of the mainframe.
Name
Type
< supply > enum P5, P12, N12, P24,
N24, N5PT2, N2,
TOTal
Type
Range
float
These maximums can be reset by powering the mainframe off and then
powering it back on.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-221
"Settings Conflict"
-224
"Illegal Parameter"
A *RST has no effect on the HISTory subsystem.
HISTory:POWer:MAXimum?, HISTory:POWer:[HISTogram]?
<supply> returns a single floating point number for
Range
Default
none
Selects the power supply for the current
maximum request. P5 is positive 5 Vdc
supply, N12 is negative 12 Vdc supply,
etc.
Maximum power (in watts) for specified power supply.
The power supply doesn't support current
measurements.
The <supply> parameter was not correct.
Description
Description
SYSTem:ERRor?
Probable Cause
query.
Chapter 3