Status:questionable:umcounter:tinterval <Time> Status:questionable:umcounter:tinterval - HP E8402A User And Service Manual

Vxi c-size mainframe
Table of Contents

Advertisement

STATus:QUEStionable:UMCounter:TINTerval < time >
STATus:QUEStionable:UMCounter:TINTerval?
Parameters
Returned Data
Comment
Reset Condition
Related
Commands
116 Programming the Enhanced Monitor
The
STATus:QUEStionable:UMCounter:TINTerval
value of the user maintenance counter. The user maintenance counter is a countdown
timer that counts hours down to zero. The time interval is the value from which the
UMCounter will start. This value is stored in non-volatile memory immediately.
The S
TATus:QUEStionable:UMCounter:TINTerval?
representing the time interval value of the user maintenance counter.
Name
Type
Range
< time >
int16
1 to
65535,MIN,
MAX, DEF
Type
Range
int16
The value 65535 means that the counter is disabled. It is shipped from the
factory disabled. Set the time interval (
the counter (
STAT:QUES:UMC:TRES
To disable the counter, send the command
STATus:QUEStionable:UMCounter:TINTerval DEF
The MINimum value is 1, MAXimum value is 65534, DEFault value is 65535.
*RST does not affect the UM Counter.
STATus:QUEStionable:UMCounter:TRESet,
STATus:QUEStionable:UMCounter:TREMaining?
Default
65535
number of hours the UMCounter will count
down from.
User Maintenance Counter
STAT:QUES:UMC:TINT
) to enable the user maintenance counter.
command sets the time interval
command returns an integer
Description
Description
) and then reset
.
Chapter 3