The RS-232 interface on the rear panel of the Enhanced Monitor mainframes
(HP E8402 and E8404) can be used to control the Enhanced Monitor from a
computer or a terminal.
Refer to Chapter 3 for RS-232 programming information. The
SYSTem:COMMunicate:SERial ...
configuration of the Enhanced Monitor's serial interface. Serial
communication commands take effect after the end of the program message
containing the command(s).
Default RS-232 parameters are:
Baud: 9600
Bits: 8
Parity: None
Stop bits: 1
DTR/RTS: On
Pace: XON
Echo: On
ERES: On
Line buffer: On
Note
If you use the Enhanced Monitor RS-232 interface (located on the
back of the mainframe) while the mainframe is in the standby mode,
you must supply an external +5Vdc to the +5VEXT connector
(located near the RS-232 interface). If you use the RS-232 interface
while the mainframe is powered on, you do not need to provide the
external +5Vdc.
The External +5V supply is for powering the Enhanced Monitor (including
the RS-232 interface) while the mainframe is powered down (standby
mode). This requires a stable 5 volt supply capable of 1.5A amps maximum
(500mA typical, refer to specifications in Appendix A). Refer to Figure 1-4.
commands set and/or modify the