Status:questionable:blower:speed? <Blower>[,Min|Max] - HP E8402A User And Service Manual

Vxi c-size mainframe
Table of Contents

Advertisement

STATus:QUEStionable:BLOWer:SPEed? <blower>[,MIN|MAX]
Parameters
Returned Data
Comments
Reset Condition
Related
Commands
Chapter 3
The
STATus:QUEStionable:BLOWer:SPEed?
the present fan speed in revolutions per minute (RPM).
Name
Type
Range
< blower >
enum. BLOWer1,
BLOWer2,
BLOWer3
optional
enum
MIN
parameter
MAX
Type
Range
int16
0 - 3500
All fans operate at the same levels as a percentage of their full scale
(
STAT:QUES:BLOW:LEV?
operate at higher speeds than the mainframe cooling fan. The measurement
returned is the last measurement made on the fans. This should be within the
last 2 seconds. If the power of the mainframe is off, this query will return the
last measurement made before power turned off.
*RST does not affect the fan speed directly. If the fans were on full speed because
the S
YSTem:BLOWer:STATE FULL
that state to VARiable
STATus:QUEStionable:BLOWer:LEVel?
query returns an integer representing
Default
none
Indicates which fan speed is being queried.
BLOWer1 is the main cooling fan, BLOWer2 is
the Power Supply cooling fan, BLOWer3 is a
second Power Supply cooling fan on the HP
E8404A mainframes.
When used, returns maximum or minimum
RPM for the present fan level.
Description
The last measured fan speed, given in RPM.
returns that level). However, the power supply fans
command had been executed, then *RST clears