STATus:SCONdition?

Returned Data
Type
uint32 0 to
uint32 0 to 31
124 Programming the Enhanced Monitor
The
STATus:SCONdition?
condition in the Enhanced Monitor. The integers are the sum of the decimal values
of the bits that are set. A returned value of 0 in both integers indicates that there are
no failures. A value of 1 (one) in a bit indicates that the corresponding present
condition has failed. Unused bits return 0. Reading the condition does not clear the
failure.
Range
Default
none
2147483647
none
a. Fan is operating out of expected range for the present level.
query returns two integers indicating the pass/fail
Decimal
Weight
1
Bit 0: Maintenance Counter Expired
2
Bit 1: +5V Power Supply Voltage out of limit
4
Bit 2: +12V Power Supply Voltage out of limit
8
Bit 3: -12V Power Supply Voltage out of limit
16
Bit 4: +24V Power Supply Voltage out of limit
32
Bit 5: -24V Power Supply Voltage out of limit
64
Bit 6: -5.2V Power Supply Voltage out of limit
128
Bit 7: -2V Power Supply Voltage out of limit
256
Bit 8: +5V STby Power Supply Voltage out of limit
512
Bit 9: +5V Power Supply Current out of limit
1024
Bit 10: +12V Power Supply Current out of limit
2048
Bit 11: -12V Power Supply Current out of limit
4096
Bit 12: +24V Power Supply Current out of limit
8192
Bit 13: -24V Power Supply Current out of limit
16384
Bit 14: -5.2V Power Supply Current out of limit
32768
Bit 15: -2V Power Supply Current out of limit
65536
Bit 16: Slot 0 Temperature over limit
131072
Bit 17: Slot 1 Temperature over limit
262144
Bit 18: Slot 2 Temperature over limit
524288
Bit 19: Slot 3 Temperature over limit
1048576
Bit 20: Slot 4 Temperature over limit
2097152
Bit 21: Slot 5 Temperature over limit
4194304
Bit 22: Slot 6 Temperature over limit
8388608
Bit 23: Slot 7 Temperature over limit
16777216
Bit 24: Slot 8 Temperature over limit
33554432
Bit 25: Slot 9 Temperature over limit
67108864
Bit 26: Slot 10 Temperature over limit
134217728
Bit 27: Slot 11 Temperature over limit
268435456
Bit 28: Slot 12 Temperature over limit
536870912
Bit 29: Ambient Temperature over limit
1073741824
Bit 30 Power Supply Temperature over limit
2147483648
Bit 31: Not Used; set to 0.
1
Bit 0: Total power
2
Bit 1: Blower 1 (main cooling fan)
Bit 2: Blower 2 (1 st Power Supply cooling fan) a
4
Bit 3: Blower 3 (2 nd Power Supply cooling fan) a
8
16
Bit 4: 5V External Voltage
32
Bit 5 through Bit 31: Not used; set to 0.
Description
a
Chapter 3