STATus:QUEStionable:TEMPerature:LIMit <slot>,<value1>[,<value2>[,<value3>]]
STATus:QUEStionable:TEMPerature:LIMit? <slot>[,MIN|MAX]
Parameters
Returned Data
Comments
114 Programming the Enhanced Monitor
The
STATus:QUEStionable:TEMPerature:LIMit
temperatures in the mainframe. If the limit is exceeded, a warning is issued. These
values are stored in non-volatile memory.
The
STATus:QUEStionable:TEMPerature:LIMit?
representing the limit for temperatures in the specified mainframe slot.
Name
Type
< slot >
enum. OUT0...OUT12,
DELTa0...DELTa12
ALL
AMBient
< value1 > int16
0 to 75, MAX, MIN,
DEF
< value2 > int16
0 to 55, MAX, MIN,
DEF
< value3 > int16
0 to 65, MAX, MIN,
DEF
Type
Range
int16
There are two limits for each slot in the mainframe: the absolute limit (default
at 65) and the delta limit (default at 15 above the ambient air reading).
Normally, the delta limit predominates the actual limit point—where a warning
will sound. But as the ambient (mainframe intake air) increases, the delta limit
may pass over the absolute limit. The absolute limit will then dominate the
warning trip point. The actual limit point can be queried with the MAX
optional parameter on the
This command determines if <value> is within the valid range for the specified
<slot>; if it is not, the limit is set to MAX. No error is generated. Use MIN and
MAX parameters with the query to determine the maximum and minimum
temperature limits.
If having the temperature warning trip point tied to the ambient air is
undesirable, set the delta limits to 55. This effectively moves them out of
range.
Remember to execute a
This will move them into non-volatile memory.
Refer to "Setting Enhanced Monitor Limits" on page 37 for information on
selecting and using temperature limits.
Range
Default
65
Absolute limit for slots 0 - slot 12
15
Delta limit above ambient for slots0 - slot
12
sets every temperature limit.
55
Absolute limit for air entering mainframe.
Temperature limit. With ALL it is the
absolute slot temperature limit.
Delta slot temperature limit. Use only with
ALL.
Ambient temperature limit. Use only with
ALL.
Temperature limit for < slot >.
STAT:QUES:TEMP:LEV?
SYSTem:NVSave
°
sets the limit, in
C, for
query returns an integer
Description
Description
query.
after setting all the limits you desire.
Chapter 3