HISTory:TEMPerature:MAXimum? < slot >
HISTory:TEMPerature:MINimum? < slot >
Parameters
Returned Data
Comments
Error Conditions
Reset Condition
Related
Commands
78 Programming the Enhanced Monitor
HISTory:TEMPerature:MAXimum?
the maximum temperature measured by the Enhanced Monitor for the <slot> since
either manufacture of the mainframe or the most recent HIST:RES:TEMP
HIST:RES:ALL.
HISTory:TEMPerature:MINimum?
the minimum temperature measured by the Enhanced Monitor for the <slot> since
either manufacture of the mainframe or the most recent HIST:RES:TEMP.
Name
Type
< slot >
enum.
OUT0...OUT12,
DELTa0...DELTa12,
AMBient
Type
Range
float
The maximum/minimum temperature tracking can be zeroed by the
commands.
HISTory:RESet
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-109
"Missing parameter"
-224
"Illegal Parameter"
-311
"Memory error"
A *RST has no effect on the HISTory subsystem.
HISTory:TEMPerature:[HISTogram]?, HISTory:RESet:TEMPerature,
HISTory:RESet[:ALL]
<slot> returns a single floating point number for
<slot> returns a single floating point number for
Range
Default
none
minimum or maximum temperature
You must include the < event index > parameter.
the <slot> parameter was not correct.
the historical data's memory is not available.
Description
Exhaust temperature for slots 0 - slot
12
Temperature change above the
ambient temperature for slots0 - slot
12
Entrance air temperature
Description
SYSTem:ERRor?
Probable Cause
or
query.