Connecting The Mainframe To A Permanent Earth Ground - HP E8402A User And Service Manual

Vxi c-size mainframe
Table of Contents

Advertisement

Connecting the
Mainframe to a
Permanent Earth
Ground
Figure 1-1. Connecting an HP E840xA Mainframe to a Permanent Earth Ground
WARNING
AVERTISSEMENT
Chapter 1
The mainframe must be connected to a permanent earth ground for line
frequencies greater than 66 Hz. This connection is made on the back of the
mainframe:
1. Connect a 16 AWG (1.3 mm or larger) wire to the PEM nut shown in
Figure 1-1. The wire must be green with a yellow stripe, or bare (no
insulation). Use a m4 x 10 screw, grounding lug, and toothed washers
(or toothed lug) as shown in the Figure.
2. Attach the other end of the wire to a permanent earth ground using
toothed washers or a toothed lug.
TOOTHED
WASHERS
PERMANENT EARTH
GROUND CONNECTION
(HP E8404 shown, HP E8402 is similar)
For protection from electrical shock when operating at
frequencies greater than 66 Hz, connect the chassis ground
terminal to permanent earth ground.
Risque de Choch èlectrique. Si la frèquence du secteur est
supèrieure à 66 Hz, relier la borne de masse du chassis à une
prise de terre fixe.
GREEN WIRE
WITH YELLOW
WIRE LUG
STRIPE
BACK COVER
LOCATION
Getting Started 19

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8404a

Table of Contents