System:communicate:vxi:address? <Address - HP E8402A User And Service Manual

Vxi c-size mainframe
Table of Contents

Advertisement

SYSTem:COMMunicate:VXI:ADDRess? < address >
Parameters
Returned Data
Comments
Reset Condition
142 Programming the Enhanced Monitor
SYSTem:COMMunication:VXI:ADDRess
the enhanced monitor device. This address is set by a switch on the back of the
mainframe.
Name
Type
< address >
enum.
Type
Range
int16
1 through 254, or 32767 224
The MIN parameter returns the lowest possible address, 1 but does not change
the actual VXI logical address used with the Enhanced Monitor. Similarly, the
MAX parameter returns the largest possible VXI address, 254 but does change
the address used. DEF returns the factory default address of 224.
This query will return the VXI logical address unless the mainframe's power is
off and the enhanced monitor device cannot read it. It then returns the value
32567.
*RST has no effect on the VXI logical address.
returns the current VXI logical address of
Range of Values
MIN|MAX|DEF
Default
VXI logical address
Default
none
Description