CALibration:VALue:TEMPerature <value>
CALibration:VALue:TEMPerature?
Parameters
Returned Data
Comments
Error Conditions
Reset Condition
Related
Commands
Chapter 4
CALibration:VALue:TEMPerature
externally measured temperature measured at a sensor inside the mainframe in a
constant temperature environment with no modules installed in the mainframe. The
value will be used to calibrate each temperature sensor of the enhanced monitor.
If a value of -1 is provided, the calibration command will calculate a calibration
temperature value from the average of all slot cavity sensors. Since the temperature
sensors are absolute temperature devices, this is adequate for most calibration
purposes.
CALibration:VALue:TEMPerature? returns the externally measured temperature
stored by the CALibration:VALue:TEMPerature command. If the value of -1 was
stored by the CAL:VAL:TEMP command and a CAL:TEMP? or CAL? was
performed, then CAL:VAL:TEMP? will return the averaged value calculated by the
calibration routine.
Name
Type
Range
< value >
float32
-1, 0-55
Type
Range
Default
float32
-1, 0-55
-1
The sensor used for this measurement should be located in the center of the
mainframe's slot cavity. It should have an accuracy of 0.5 °C. If you want the
calibration routine to calculate a temperature calibration value you MUST
send a CAL:VAL:TEMP -1 before every calibration command.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using SYSTem:ERRor? query.
Number
-221
"Settings Conflict"
Calibration is not affected by *RST.
CALibration:ALL?
<value> provides the enhanced monitor with an
Default
-1
Provides the external measurement of ambient
temperature in
Enhanced Monitor to use the average value of
all 39 slot exhaust temperature sensors as the
calibration value.
Returns the stored external measurement of ambient
°
temperature in
Message
The mainframe's power is off.
Calibrating and Verifying Performance
Description
°
C. issuing a -1 forces the
Description
C.
Probable Cause
201