Calibration
12.3
Calibrating the Fluoride Sensor
12.3.1
Selection of the calibration process for fluoride
92
Continuous display ➨
CAL ORP
Offset
Last calibration
CAL setup
Calibration process
continue with <CAL>
Fig. 56: Data input, ORP sensor calibration (CAL)
1.
Continue with
CAL ORP
Offset
Accept with <CAL>
Fig. 57: Adjusting the [Offset]
2.
Press
and use the four arrow keys to adjust the mV value
of the buffer you are using
3.
Press
4.
Transfer the result of the calibration into the controller
memory by pressing
ð The controller operates with the calibration results.
To calibrate the controller there are two available calibration pro‐
cesses:
1 point
n
2 point
n
0.0 mV
21.05.2013 14:59:56
Data input offset
0.1 mV
A1032
A1033
Need help?
Do you have a question about the DULCOMETER diaLog DACb and is the answer not in the manual?