Setting The [Ma Outputs] - ProMinent DULCOMETER diaLog DACb Assembly And Operating Instructions Manual

Multi-parameter controller
Hide thumbs Also See for DULCOMETER diaLog DACb:

Advertisement

[mA outputs]
Setting the
19
Setting the
mA outputs
mA output 1
mA output 2
mA output 3
Fig. 110: Setting the [mA outputs].
146
[mA outputs]
User qualification: trained user,
n
tion' on page 16
Continuous display ➨
[mA outputs]
[Channel 2] and [Channel 3]
Settings for
The 2-channel version of the controller has 2 mA outputs, and
n
the 3-channel version has 3 mA outputs. The descriptions of
[Channel 1] correspondingly apply to the settings of
[Channel 2] and [Channel 3] . The procedure for setting the
respective mA output channels is identical, however the
parameters to be set may differ. Your attention is drawn to the
differences, which are also described.
CAUTION!
Destruction of the monitors
Only connect passive monitors to the mA outputs.
For example, if the mA outputs are connected to a
PLC, then a 4-wire connection type must be
selected on the PLC. The 2-wire connection type
can result in incorrect operation and, possibly, the
destruction of the monitors.
In its basic version, the controller has 2 active mA outputs,
meaning that the mA outputs actively supply an output current,
without an external supply voltage being provided. The mA outputs
are galvanically isolated.
Response with
response of the mA outputs when
Optional mA outputs
The menu items for the optional mA outputs have the same
n
setting options as the
description is not provided.
Ä Chapter 2.4 'User qualifica‐
[mA outputs] ➨
or
[Pause Hold] : [Pause Hold] determines the
[Pause Hold] is active.
[mA output 1] menu item. A separate
8.1

Advertisement

loading
Need help?

Need help?

Do you have a question about the DULCOMETER diaLog DACb and is the answer not in the manual?

Questions and answers

Andrew Healey
March 11, 2025

I need to recalibrate my prominent DiaLog DACb are there written instructions I can use

1 comments:
Mr. Anderson
May 14, 2025

To recalibrate the ProMinent diaLog DACb, follow these steps:

1. Access the calibration function via the controller menu.
2. Use the arrow keys to navigate through the options.
3. Select the appropriate channel and parameter (e.g., Chlorine).
4. Adjust the slope (e.g., 5.99 mA/ppm) and zero point (e.g., 4.00 mA) as needed.
5. Confirm and save the new calibration values.

Ensure the calibration log book is active to store the changes.

This answer is automatically generated