Calibration
Calibrating the Fluoride
9.3
Sensor
Selection of the calibration
9.3.1
process for fluoride
To calibrate the controller there are two
available calibration processes:
1 point
n
2 point
n
Calibration process selection
1.
Continuous display ➨
CAL F -
1 ppm =
185.0 mV
Slope
-59.16 mV/dec
100 %
Single point calibration
Two point calibration
Fig. 50: Calibration menu [Fluoride]
ð The calibration menu is dis‐
played.
2.
Using the arrow keys select the
desired menu item. Press the
key
ð You can now start the selected
calibration process.
80
9.3.2
16:51:18
11/11/2011
11:11:11
11/11/2011
Material required for calibration of fluoride
sensors:
A1037
n
2-point fluoride sensor cali‐
bration (CAL)
Correct sensor operation
Correct measuring and metering
–
is only possible if the sensor is
working perfectly
Observe the sensor operating
–
instructions
The carrying out of a 2-point cali‐
–
bration is strongly recommended
and is to be preferred to other
methods
For calibration the sensor must
–
be removed and refitted in the in-
line probe housing. To do this,
observe the operating instructions
of your in-line probe housing
Two test containers with calibrating
solution