2-Point Calibration Of The Ph Sensor (Cal) - ProMinent DULCOMETER diaLog DACb Assembly And Operating Instructions Manual

Multi-parameter controller
Hide thumbs Also See for DULCOMETER diaLog DACb:

Advertisement

12.1.2

2-Point Calibration of the pH Sensor (CAL)

2.
Press
CAL pH
Calibration process
Buffer detection
Buffer manufacturer
Buffer value 1
Buffer value 2
Buffer temperature
Fig. 41: Selecting the calibration process
ð The menu for selecting the calibration process appears.
3.
Use the arrow keys to select the required menu item and
press
ð The input window appears and you can make the neces‐
sary settings for your process
4.
Use the arrow keys to select the calibration process and
press
5.
Continue with
ð You can now start your chosen calibration process.
Correct sensor operation
Correct measuring and metering is only pos‐
sible if the sensor is working perfectly.
Refer to the sensor's operating instructions.
2-point calibration is strongly recommended
and is preferable to other methods.
The sensor needs to be removed and refitted
in the in-line probe housing for calibration.
Refer to the operating instructions for your in-
line probe housing.
Calibration
2 point
requirement
ProMinent
pH 7
pH 4
Off
A1025