Technical data and measuring ranges
Designation
Maximum operating height:
Ambient conditions:
Ambient temperature:
Low voltage cables:
Degree of protection:
Material:
Dimensions:
Weight:
166
Technical data
maximum 2000 metres above sea level (m.a.s.l.)
Use indoors or with protective enclosure
Maximum 95% relative air humidity (non-condensing)
Ambient temperature - 20 ... +50 °C
Copper low voltage cables are needed with a temperature
resistance of ≧ 70 °C., min. AWG 18 or min. 0.75 mm² at the
ambient temperatures
Isolated wires, cables and other electrical conductors are
needed, which need to be flame-resistant. It is assumed that
wires with rated data in accordance with UL 2556 VW-1 or
similar normative regulations meet this requirement.
Installation on a surface: IP66/IP67
Control cabinet installation: IP54
based on NEMA 4X Indoor
Housing PC with flameproofing equipment
250 x 220 x 122 mm (WxHxD)
net 2.1 kg
The technical data on the module: 2x conductive conductivity/
temperature sensors, part number 734223, see
'Module: 2x conductive conductivity/temperature sensors. Part
number 734223' on page 51