[Relays]
Setting the
17.1.1
Function description [Off]
17.1.2
Functional description of
17.1.3
Function description
142
Changeable scope of the menus
The number of adjustable parameters may differ
depending on the type and scope of the [Function]
selected. The controller provides you with the pos‐
sible adjustable parameters. Use the
to select them and confirm with
adjustment ranges are specified by the controller.
Relay 1
Function
Function
Cycle time
Min. time
Assignment
Fig. 106: Possible adjustable parameters with [Function] include,
for example, [Control variable]
[Off] , the relay does not accept any functions or
If the setting is
allow any actions.
[Relay timer]
[Relay timer] is a real-time timer based on relay 1 and 2. The
The
[ Relay timer] lets you perform recurring weekday and time-
dependent metering processes. Timer 1 is for relay 1 / Timer 2 is
for relay 2.
[Limit 1] or [Limit 2]
[Relay 1] and/or [Relay 2] can be operated as limit value relays.
The limit values can be set in the menu
[Limit values]' on page 131 .
Limit value relay used as an actuator
Extended functions
The limit value relays can also be defined in
–
such a way that they react like an actuator. If,
for example, a limit value relay is activated,
then it is deactivated if the pause contact is
closed and for a subsequent delay period t
t
> 0 min is set).
d
or
key
. The possible
6.1.1
Control variable
Increase value
10s
1s
Channel 1
Ä Chapter 14 'Setting the
(if
d
A1071
Need help?
Do you have a question about the DULCOMETER diaLog DACb and is the answer not in the manual?
Questions and answers
I need to recalibrate my prominent DiaLog DACb are there written instructions I can use
To recalibrate the ProMinent diaLog DACb, follow these steps:
1. Access the calibration function via the controller menu.
2. Use the arrow keys to navigate through the options.
3. Select the appropriate channel and parameter (e.g., Chlorine).
4. Adjust the slope (e.g., 5.99 mA/ppm) and zero point (e.g., 4.00 mA) as needed.
5. Confirm and save the new calibration values.
Ensure the calibration log book is active to store the changes.
This answer is automatically generated