Section 5
DETAILED SETTINGS

Changing Display Scales

Change the display scale when you want to display a value different on the main display to the
actual measured value.
Place the sensing object in position and make the settings for either one-point or two-point
scaling.
The scaling set here is reflected in the display only. The output does not change.
The minimum display value is − 19,999 and the maximum is 59,999. If the measurement result is less than
the minimum display value after scaling is executed, the display will be − 19,999. If the measurement result is
greater than the maximum, the display value will be 59,999.
Scaling Application Example
To display distance from Sensor Head
Sensor
Head
Actual
measured
value:
1.0 mm
Sensing
object
To display the height of the sensing object
Sensor
Head
Sensing
object
ZX-E
90
Operation Manual
Purpose
Value to display:
2.5 mm
Actual measured value:
1.0 mm
Value to display:
1.5 mm
Settings Method
Offset the display value.
Display value (mm)
Offset the display value and invert the relation-
ship between display and reference values.
Display value (mm)
Actual measured
value (mm)
One-point scaling
with no inverted
display values
p. 92
Actual measured
value (mm)
One-point scaling
with inverted dis-
play values
p. 92