Section 5
DETAILED SETTINGS
Selection
PP-H (Peak-to-peak
hold)
S-H (Sample hold)
AVE-H (Average hold)
Moving to FUN mode and HOLD
1.
Set the mode switch to FUN.
2.
Use the LEFT and RIGHT Keys to display
HOLD on the main display.
Selecting Hold Condition
3.
Press the UP or DOWN Key.
The sub-display will flash.
4.
Use the UP and DOWN Keys to select the
hold condition.
5.
Press the ENT Key to confirm the selection.
The setting will be registered.
ZX-E
82
Operation Manual
Holds the difference between the maximum and minimum values. This option is
selected mainly when detecting vibration. The output changes at the end of the
sampling period and is held until the end of the next sampling period.
Maximum
value
Current meas-
ured value
Sampling period
Holds the measured value at the start of the sampling period. The output
changes at the start of the sampling period and is held until the start of the next
sampling period.
Current meas-
ured value
Sampling period
Holds the average measured value during the sampling period. The output
changes at the end of the sampling period and is held until the end of the next
sampling period.
Current meas-
ured value
Sampling period
Details
Output (difference between
Minimum
maximum and minimum)
value
Output
Output (average)