Setting Number Of Samples To Average - Omron ZX-E Operation Manual

Zx-e smart sensors series inductive displacement type
Hide thumbs Also See for ZX-E:
Table of Contents

Advertisement

Section 5
DETAILED SETTINGS

Setting Number of Samples to Average

The number of samples to average is the number of data points used to average data
measured by the Sensor. The average value will be output.
Use the number of samples to average function to ignore sudden variations in measured
values. If the number of samples is increased, however, the response time of the judgement
outputs and linear output will be increased.
Moving to FUN mode and AVE
1.
Set the mode switch to FUN.
2.
Use the LEFT and RIGHT Keys to display
AVE on the main display.
Selecting Number of Samples to Average
3.
Press the UP or DOWN Key.
The sub-display will flash.
4.
Use the UP and DOWN Keys to select the
number of samples to average.
5.
Press the ENT Key to confirm the selection.
The setting will be registered.
ZX-E
80
Operation Manual
Selection for No. of
samples to average
1
0.3 ms
2
0.5 ms
4
0.8 ms
8
1.5 ms
16
2.5 ms
32
5 ms
64 (default)
10 ms
128
20 ms
256
40 ms
512
75 ms
1024
150 ms
2048
300 ms
4096
600 ms
Response time