Correcting Linear Output Values
Discrepancies may occur between the linear output current (or voltage) values set on
the Amplifier Unit and the actual current (or voltage) values measured due to the condi-
tions for the connected external device or other factors. The linear output correction
function can be used to correct this discrepancy.
The output values are corrected by entering the correction value for the current or volt-
age values for any two points.
Set the monitor focus function and select either current or voltage output beforehand.
p. 103
This section uses a current output as an example. Change the values in this example
for voltage output as necessary.
1.
Connect the linear output to an external ammeter.
Moving to FUN Mode and SPCL
2.
Turn ON the power supply and set the mode
switch to FUN.
3.
Use the LEFT and RIGHT Keys to display
SPCL on the main display.
Moving to LEFT-ADJ
4.
Press the UP or DOWN Key.
The sub-display will flash.
5.
Use the UP or DOWN Keys to display SET or
ALL.
Section 5
DETAILED SETTINGS
ZX-E
Operation Manual
107