Purpose
To correct display values to match actual dis-
tances
Sensor Head
Actual distance: 1.0 mm
Value to display: 0.8 mm
Sensing
object
Slide gauge
Actual distance: 0.2 mm
Value to display: 0.3 mm
Sensing
object
The settings listed below return to the default settings when scaling is set.
Make the settings for these items after scaling settings have been completed.
Self-trigger level, p. 83
Output Settings (Monitor Focus), p. 103
Using the Zero Reset Function, p. 127
Finding Thicknesses, p. 118
Settings Method
Offset the display value and change the dis-
play value range.
Display value (mm)
Actual measured
value (mm)
Section 5
DETAILED SETTINGS
Two-point scaling
p. 95