Section 5
DETAILED SETTINGS
Setting the Second Point
2.
Place the sensing object in the position for which the display is to be changed (the sec-
ond point).
The sensing object must be set at a distance at
least 1% of the rated measurement distance
away from the first point and also at a distance
within the measurement distance range.
3.
Press one of the Cursor Keys.
The current measured value will be displayed on the
main display. The leftmost digit of the sub-display will
flash.
4.
Use the Cursor Keys to set the offset for the
measured value.
The position of the decimal point can be changed using
the next steps.
5.
Press the ENT Key to confirm the settings.
The decimal point will flash.
6.
Use the LEFT and RIGHT Keys to move the
decimal point.
7.
Press the ENT Key to confirm this setting.
Confirming the Completion of the Scaling Settings
If scaling has been completed correctly, the display will
be OK.
If scaling was unable to be completed, the display will be
NG. Check the following points and then execute scaling
again.
Is the sensing object within the measurement distance?
Are the two points separated by at least 1% of the rated
measurement distance?
ZX-E
96
Operation Manual
Move between digits.
Increment and decre-
ment the numeric value.