Section 5
DETAILED SETTINGS
Output Settings for Non-measurement
The linear output method for when a reset is input can be set.
Selection
KEEP (default)
CLAMP
For Hold Measurements
Even if KEEP is set, the output before the first hold value is obtained will be the same as CLAMP.
Moving to FUN Mode and SPCL
1.
Set the mode switch to FUN.
2.
Use the LEFT and RIGHT Keys to display
SPCL on the main display.
Moving to RESET
3.
Press the UP or DOWN Key.
The sub-display will flash.
4.
Use the UP and DOWN Keys to display ETC
or ALL.
5.
Press the ENT Key.
ZX-E
110
Operation Manual
Judgement outputs
The status immediately before measurement is stopped is held and output.
All OFF
Outputs
Linear output
Outputs the set CLAMP value.
The following options are available.
• For current output: 3 to 21 mA or maximum
(approx. 23 mA)
• For voltage output: − 5 to 5 V or maximum
(approx. 5.5 V)