Setting Thicknesses
7.
Use the LEFT and RIGHT Keys to display
THICK on the main display.
8.
Press the UP or DOWN Key.
The leftmost digit of the sub-display will flash.
9.
Set the thickness of the standard sensing
object.
Enter the actual dimensions of the reference object.
The position of the decimal point can be changed using
the next steps.
10.
Press the ENT Key to confirm the setting.
The numeric values will be confirmed and the decimal
point will flash.
11.
Use the LEFT and RIGHT Keys to move the
decimal point.
12.
Press the ENT Key to confirm the position of
the decimal point.
The scaling values (A and B) are registered for both
Amplifier Units.
Countermeasures for Setting Errors
If the display shows E-THK, the standard sensing object
is outside the measurement distance. (The ENABLE indi-
cator will not be lit.)
Adjust the position of the reference object until the
ENABLE indicator is lit on both Amplifier Units and exe-
cute the measurement again.
Thickness
Section 6
AUXILIARY FUNCTIONS