UTG9000T Series Function Abitrary Waveform Generator
2) Tap Duty tab, pop-out virtual numeric keyboard to enter 70%.
3) Tap word Duty again to step through Duty/PWidth.
Figure 2-7 Square Settings

2.3.5 Pulse Wave Settings

The duty ratio of pulse wave presents the time quantum between with threshold value of rising edge 50% decrease to the
next falling edge 50% (assuming that the waveform is not inverse.)
Users can make parameter settings to this instrument, then it can output the adjustable pulse wave with pulse width and
edge time. The duty cycle default value is 50% of the pulse wave, rising/falling edge time 1us.
The step to set period 2ms, amplitude 1.5Vpp, DC offset voltage 0V, duty ratio 25 %( limited by the lower pulse wave width
2.4ns), rising/falling edge time 200us:
1)Tap Continue tab to select Pulse wave mode, pop-out numerice keyboard to enter 1.5Vpp.
2)Tap Duty tab, pop-out the virtual numeric keyboard to enter 25%.
3)Tap REdge tab, pop-out the virtual numeric keyboard to enter 200us, the same way to set the FEdge.
Figure 2-8 Pulse Wave Settings

2.3.6 DC Voltage Settings

The default value is 0V of the DC voltage.
The step to set DC offset voltage to 3V: