Set Square Wave; Set Pulse Wave - UNI-T UTG2122B User Manual

Function/arbitrary waveform generator
Hide thumbs Also See for UTG2122B:
Table of Contents

Advertisement

become high level (MAX) and low level (MIN).

2.4.4 Set square wave

Default duty ratio: 50% limited by minimum pulse width 6.5ns. Steps to set a square wave with 1Hz,
1.5Vpp, DC offset 0V, duty ratio 70%:
1.
Press Menu → Wave → Type → Square → Param
2.
Enter required numerical value and select the unit.

2.4.5 Set pulse wave

Example: period=2ms, amplitude=1.5Vpp, offset=0V, duty=25%, rising time=200us and falling
time=200us:
1.
Press Menu → Wave → Type → Pulse → Param , and then press Freq to switch to
Period. Enter required number value and select the unit. When entering duty ratio value, there is a
label at the bottom of display, and select 25%.
2.
Press Param to enter sub menu to set falling edge time, then press Fall to enter required
14
UTG2000B Series