UNI-T UTG9504T User Manual page 15

Function/ arbitrary waveform generator
Table of Contents

Advertisement

UTG9000T Series Function Abitrary Waveform Generator
②.External 10MHz intput terminal
Establish synchronization of multiple function/arbitrary waveform generators or synchronization with external 10 MHz clock
signal. When clock source of the instrument is external, external 10MHz input terminal receives an external 10MHz clock
signal.
③.Internal 10MHz output terminal
Establish synchronous or external clock signal with reference frequency of 10 MHz for multiple function/arbitrary waveform
generators. When clock source of the instrument is internal, internal 10MHz output terminal outputs an internal 10MHz clock
signal.
④. Frequency Counter Interface
Input signal throught the interdace when using frequency counter.
⑤.External Digital Moudlation Interface
In case of modulation of ASK, FSK, PSK or OSK signal, if modulation source is external, input modulation signal through
external digital modulation interface (TTL level). The corresponding output amplitude, frequency and phase are determined
by signal level of external digital modulation interface. If trigger source of frequency sweep is external, receive a TTL pulse
with designated polarity through external digital modulation interface. This pulse can start scanning. If burst mode is gated.
Trigger source of N period and wireless trigger source are external, input gated signal through the external modulation
interface. This pulse string can output a designated cycle number of pulse string.
⑥.External Analog Modulation Output Terminal
In case of AM, FM, PM, DSB-AM, SUM or PWM signal, if modulation is external, input signal through through external analog
modulation. The corresponding modulation of depth, frequency deviation, phase deviation or duty ratio deviation is
controlled by ±5V signal level of the external analog modulation input terminal.
⑦.USB Interface
Connect with the upper computer software through USB interface to achieve the contr of the instrument by computer.
⑧.LAN Port
The instrument can connect with LAN by LAN port, to achieve remote control.
⑨.AC Power Input Terminal :AC power :100~240V,45~440Hz, power fuse: 250V,T2A.
⑪.Case Locker
⑩.Main Power Switch:Power on in "I"position; Power off in "O"position ( The front panel ON/OFF button is unable to use.)
Open the case locker to activate the function of anti-theft.