UTG9000T Series Function Abitrary Waveform Generator
Figure 2-5 Amplitude Settings
2.3.3 DC Offset Voltage Settings
The instrument output DC offset voltage of a sine wave amplitude is 0V (default setting) when activating the instrument.
The step to set the DC offset voltage to-150mV:
1)Tap Continue tab to select Sine
2)Tap Offset tab, pop-out the virtual numeric keyboard to enter -150mV
3)Tap word Offset, Amplitude and Offset tab becomes High (maximum)/Low (minimum) leve. This method is convenient to
set the signal limits of digital applications
Figure 2-6 DC Offset Voltage Settings
2.3.4 Squre Wave Settings
The duty ratio of square wave presents the time quantum of the square wave at a high leve of each cycling
(assuming that the waveform is not inverse.) The duty ratio default value is 50% of the square wave.
The step to set frequency to 1kHz, amplitude 1.5Vpp, DC offset voltage 0V, duty ratio 70%:
1) Tap Continue tab to select Square wave mode, tap Amplitude tab to pop-out virtual numeric keyboard to enter
1.5Vpp.
19