Fine Adjustment of the Well-plate Handler Position

These adjustments are made by moving the Well-plate Handler arm and
gripper fingers using the Test functions of the Well-plate Handler. See
with the Agilent ChemStation"
Controller"
You cannot complete these fine adjustments until you have completed the Well-plate
N O T E
Handler Teaching described in

Adjusting the radius

We recommend that you make these adjustments using the 54-vial tray, with vials in
N O T E
positions C1 and D1. This is especially important if you intend to use the 54-vial tray in your
work.
1 Place a plate in the rack R1.
2 Use the Test functions of the Well-plate Handler to
3 Ensure that the plate detection button clears the vials and makes contact
4 Use the Test functions of the Well-plate Handler to
5 If the plate does not sit correctly on the transport position of the
48
on page 87 for details of the available test functions.
a position the gripper arm above rack R1,
b move the arm down until it encounters the plate,
with the rim of the plate.
a close the gripper fingers,
b raise the arm until the plate is clear of the rack,
c position the arm above the Automation Interface,
d lower the arm until the rack is placed on the transport position of the
Automation Interface.
Automation Interface,
a raise the plate clear of the Automation Interface,
on page 84 or
"Well-plate Handler Teaching"
Agilent 1200 Series Sample Capacity Extension User Manual
"Testing with the Hand-Held
on page 41.