Installation
Vertical Teaching
The vertical teaching calibrates two positions:
• the Automation Interface Height calibrates the height of a well-plate on the
• the Well-plate Handler Ground calibrates the height of a single well-plate in the
Automation Interface
Height
A plate must be installed in the transport position of the Automation Interface.
N O T E
Well-plate Handler
Ground
A single plate must be installed in rack R1.
N O T E
Vertical Teaching Instructions
1 Place a well-plate in the transport position of the Automation Interface.
2 Click Automation Interface Height.
3 Select the plate type from the list displayed in the dialog box.
4 Remove all well-plates from rack R1.
5 Transfer the well-plate from the Automation Interface to rack R1.
6 Click Well-plate Handler Ground.
7 Select the plate type from the list displayed in the dialog box.
8 Click Save to save the coordinates.
44
Automation Interface transport position.
rack.
Moves to position R6, and moves the arm down until it
encounters a plate.
Moves to position R1, and moves the arm down until it
encounters a plate.
Agilent 1200 Series Sample Capacity Extension User Manual