Installation
1 Click Default to set the default values.
2 Select position R1 and click OK.
3 Click the cursor down button until the arm sits just above the rack.
4 Click the cursor left and right buttons until the gripper is centralized in the
You can make a more accurate judgement of the alignment by placing a plate in the rack
N O T E
and moving the gripper arm down until it sits just above the plate.
5 Click OK to set the position.
6 Click the cursor up button until the arm is clear of the rack.
7 Select a new rack position and repeat instructions 3 – 6.
All positions must be taught, including those of the optional rack extension, if fitted.
N O T E
8 Select rotary position R6 (Automation Interface) and click OK.
9 Click Move Down until the arm sits just above the transport position.
10 Repeat instructions 3 – 6.
You can make a more accurate judgement of the alignment by placing a plate on the
N O T E
transport position of the Automation Interface and moving the gripper arm down until it sits
just above the plate.
Vertical Teaching Instructions
The two vertical positions, Automation Interface Height and Well-plate Handler
Ground, (see
1 Place a well-plate in the transport position of the Automation Interface.
2 Select Automation Interface Height.
3 Select the plate type from the displayed list and click OK to continue.
4 Remove all well-plates from rack R1.
5 Transfer the well-plate from the Automation Interface to rack R1.
6 Select Well-plate Handler Ground.
46
rack.
Figure 34
on page 45), are calibrated automatically.