De-Installation of the Automation Interface

The Automation Interface is incompatible with the temperature control of the
high performance autosampler. If you want to use the high performance
autosampler with temperature control, you must de-install the Automation
Interface.
1 Disconnect all connections to the Automation Interface
2 Remove the two lock screws holding the Automation Interface (see
3 Withdraw the Automation Interface from the high performance
4 Remove the adapter plate.
5 Replace the two rubber plugs (see
6 Install the high performance autosampler tray.
7 In the Trays group of the Injector Configuration dialog box of the Agilent
8 Re-install the plastic screen from the original cover of the high performance
9 Turn on the Agilent 1200 Series LC system including the high performance
10 Start the ChemStation SW and type the following command into the
11 Reset the high performance autosampler.
Agilent 1200 Series Sample Capacity Extension User Manual
Figure
10)
autosampler and replace the lock screws in position in the Automation
Interface.
ChemStation, ensure that a tray is configured, not the Automation Interface
(see
"Well-plate Handler Configuration"
autosampler.
autosampler. The autosampler will abort the initialization and the status
LED will turn red.
command line:
print sendmodule$ (lals, "MAMI:NMAI")
Figure
5).
on page 60).
Installation