Troubleshooting and Error Handling

Changing the Fuse(s)

Fuses burn out occasionally and must be replaced. If the power indicator light
on the front of the unit does not light after switching ON the Well-plate
Handler, first check to see whether the power cord is securely seated in a
functioning power outlet and in the Well-plate Handler power cord receptacle
at the rear of the Well-plate Handler. If the power failed while the Well-plate
Handler was ON, check that the power cord is not loose or disconnected and
that power to the outlet is functioning properly. If these checks fail to remedy
the loss of power, follow the steps listed below to replace the fuses.
Spare fuses are shipped with Well-plate Handler. If you no longer have spare
fuses, you may obtain new ones from Agilent Technologies. The Well-plate
Handler uses the following fuses:
• U.S.: 1.5 amp, 0.25 x 1.25 inch
• Metric: 1.0-amp, 5.0 x 20 mm
Make sure the power to the Well-plate Handler is OFF and that the power cord is
WA R N I N G
unplugged from the instrument or the power source before proceeding with the
following instructions.
1 Switch OFF power to the Well-plate Handler and remove the power cord
2 Using a small, flat-blade screwdriver, gently pry the cover/fuse block
3 On the fuse block, loosen the Phillips-head screw two turns and remove the
4 Remove the blown fuse: either the two 1.0-amp fuses or the single 1.5-amp
5 Making sure that the active fuses face out, slide the fuse block back into
6 Slide the cover/fuse block assembly back into the housing and snap into
7 Reconnect the power cord to the Well-plate Handler and to the wall outlet
96
from the power outlet or from the Well-plate Handler power cord
receptacle.
assembly away from its housing and remove it to access the fuses.
fuse block by sliding it up and away from the screw and pedestal.
fuse and replace with fuses of the same rating.
position and tighten the screw.
place on the rear panel of the Well-plate Handler.
and then reconnect the other cables previously disconnected.
Agilent 1200 Series Sample Capacity Extension User Manual